Avalon fetal monitor, for monitor release j.3 with software revision j.3x.xx, patient monitoring.
Hide thumbs

Advertisement

Configuring Bed Information

Monitor Settings
Bed Information Factory Default Setting
Setting Item
IP OB Server
Bed Information Configuration Implications
IP OB Server
Use the
IP OB Server
This is the IP address of any fetal monitor data acquisition host such as a data acquisition client PC, an
IntelliSpace Perinatal (formerly OB TraceVue) data acquisition server, or a surveillance and archiving
server.
If the field is left empty (set to 0.0.0.0), the fetal monitor can connect automatically to any available
data acquisition PC in the same subnet.
If the current host PC is unavailable, the fetal monitor automatically tries to find another host.
If the IP address of the host PC is entered manually, the fetal monitor can only connect to this
dedicated data acquisition PC, and dynamic hosting is not possible.

Configuring Edit Notes

Monitor Settings
Edit Notes Factory Default Settings
Setting Item
Note <1>
1
Note <2
1
Note <3
1
Note <4
1
Note <5
1
Note <6
1
Note <7
1
Note <8
Note <9>
Note <10>
Note <11>
Note <12>
Note <13>
Note <14>
:
Main Setup
>
Bed Information
Mode
C
setting to define the IP address of the OB server.
:
Main Setup
>
Edit Notes
Mode
1
C
C
C
C
C
C
C
C
1
C
1
C
1
C
1
C
1
C
1
C
Factory Default
Choice/Range
0.0.0.0
n/a
>
Edit
Factory Default
Choice/Range
Patient Repositioned
free configurable, maximum 30 characters
Vaginal Examination
free configurable, maximum 30 characters
MD Notified
free configurable, maximum 30 characters
Sitting
free configurable, maximum 30 characters
On Back
free configurable, maximum 30 characters
Left Lateral
free configurable, maximum 30 characters
Ambulating
free configurable, maximum 30 characters
Tocolytic Given
free configurable, maximum 30 characters
Membranes Ruptured
free configurable, maximum 30 characters
Amniotomy
free configurable, maximum 30 characters
Amniotic Fluid Clear
free configurable, maximum 30 characters
Amniotic Fluid Not Clear free configurable, maximum 30 characters
Oxytocin
free configurable, maximum 30 characters
Urinary Catheter
free configurable, maximum 30 characters
6 Monitor Settings
83