Configuring Bed Information Hardware Settings

Hardware Settings
Bed Information Hardware Factory Default Settings
Setting Item
Equipment Label
IP Config
IP Address
Subnet Mask
Default Gateway
1
Only accessible if
Bed Information Hardware Settings Configuration Implications
For detailed configuration implications, please refer to the latest Avalon Fetal Monitor Service Guide.
Equipment Label
The equipment label must be entered for each individual monitor. It is stored with the monitor, it is
not cloned. The equipment label must be unique for each fetal monitor in order for the fetal monitors
to be able to connect to IntelliSpace Perinatal/OB TraceVue.
IP Config
Use the
IP Config
Choices are:
BOOTP
Typically, the Avalon Fetal Monitor requests its
if a
BOOTP
is then set to a dedicated, valid address.
DHCP
The
DHCP
enables the fetal monitors to request an
network (OB TraceVue/IntelliSpace Perinatal) automatically.
DHCP Restr
In
DHCP Restr.
Manual
CAUTION
It is mandatory that
personnel. Exercise great caution during installation to avoid problems such as duplicate IP
addresses, non-matching subnet mask, etc. Documentation of all related configuration details is
mandatory and must be updated with each change to ensure network reliability, especially when
:
Main Setup
>
Bed Information
Mode
C, S
S
1
S
1
S
1
S
IP Config
is set to
setting to to change the mode to acquire the
server cannot be provided, fix parameters may be configured manually. The
support offers an alternative to
mode any true BOOTP responses are ignored.
Manual IP Address
Factory Default
Choice/Range
<empty>
n/a
BOOTP
BOOTP, DHCP, DHCP Restr. Manual
0.0.0.0
n/a
0.0.0.0
n/a
0.0.0.0
n/a
Manual
.
IP Address
from a
IP Address
.
(dynamic host configuration protocol)
BOOTP
DHCP
IP Address
(internet protocol address) from the connected
configuration is performed by experienced service
8 Hardware Settings
.
service (preferred). Only
BOOTP
IP Address