8 Hardware Settings

Configuring OBR (OB Radio) Settings

Hardware Settings
OBR Factory Default Settings
Setting Item
Revision
Frequency Band
Channel
1
The
Channel
OBR Configuration Implications
As an exception to most of the Hardware settings, OBR Settings can be cloned from one to another
fetal monitor.
Revision
This is not a setting. It displays the revision number of the OB radio.
Frequency Band
Use the
Frequency Band
WMTS
(e.g. most European countries)
ISM
T108
For detailed configuration implications, please refer to the latest Avalon fetal monitor Service Guide.
Channel
Use the
Channel
channels. The range of available channels depends on the selected
Revision
.
Depending on the hospital's existing wireless infrastructure, a number of channels might already be
occupied by other wireless applications.
For detailed instructions on setting up OBR channels in a new or existing wireless infrastructure, refer
to the Avalon Fetal Monitor Service Guide and contact a Philips service representative.
100
:
Main Setup
>
Hardware
Mode
n/a
C, S
C, S
range depends on selected
setting to configure the country-specific OBR frequency band. Choices are:
(e.g. USA, Canada)
(e.g. Japan)
setting to configure the OBR channel the monitor should use. It provides a total of 64
>
OBR
Factory Default
Choice/Range
<revision number>
n/a
Unknown
Unknown, WMTS, ISM, T108
Unknown
Unknown, 1 to 64
Frequency Band
and the OB Radio
1
Revision
.
and the OB Radio
Frequency Band