8 Hardware Settings
Configuring OBR (OB Radio) Settings
Hardware Settings
OBR Factory Default Settings
Setting Item
Revision
Frequency Band
Channel
1
The
Channel
OBR Configuration Implications
As an exception to most of the Hardware settings, OBR Settings can be cloned from one to another
fetal monitor.
Revision
This is not a setting. It displays the revision number of the OB radio.
Frequency Band
Use the
Frequency Band
•
WMTS
•
(e.g. most European countries)
ISM
•
T108
For detailed configuration implications, please refer to the latest Avalon fetal monitor Service Guide.
Channel
Use the
Channel
channels. The range of available channels depends on the selected
Revision
.
Depending on the hospital's existing wireless infrastructure, a number of channels might already be
occupied by other wireless applications.
For detailed instructions on setting up OBR channels in a new or existing wireless infrastructure, refer
to the Avalon Fetal Monitor Service Guide and contact a Philips service representative.
100
:
Main Setup
>
Hardware
Mode
n/a
C, S
C, S
range depends on selected
setting to configure the country-specific OBR frequency band. Choices are:
(e.g. USA, Canada)
(e.g. Japan)
setting to configure the OBR channel the monitor should use. It provides a total of 64
>
OBR
Factory Default
Choice/Range
<revision number>
n/a
Unknown
Unknown, WMTS, ISM, T108
Unknown
Unknown, 1 to 64
Frequency Band
and the OB Radio
1
Revision
.
and the OB Radio
Frequency Band
Comments to this Manuals