Download  Print this page

Advertisement

P
5-20
AGE
P
/ E
C
ROBLEM
VENT
ROM Hardware failure at
power on
System parameter Hardware
failure at power on
Motor re-start
SimplyGo Inhibit Battery
Charging
SimplyGo Re-enable Battery
Charging
High PCB temperature
SimplyGo High Battery
Temperature
SimplyGo Low Battery
Declared
SimplyGo Battery Depleted
Declared
SimplyGo Power Off Due To
Low Battery
SimplyGo Invalid power down
Watchdog Occurred
High O2 Sensor failure
Cooling Fan Failure
Entered the backup state
when in Night Time mode
No Flow (Kinked cannula)
Alarm
Delivery Valve Leak Failure
Delivery Valve Stuck Failure
P
C
ODE
OSSIBLE
AUSE
Main PCA
Main PCA
Compressor
High Temperature
The unit charges the
battery
High MSP Temp
High ambient
Temperature or
defective battery
Unit was not plugged in
to recharge
Defective battery
Low power level
Unplugged from the
wall AC
Unit will power off and
start running when
power is restored
Main PCA
Fan connection or Fan
Issue
No Breath detected
Kinked Cannula
Kinked internal tubing
Bad Delivery Valve
Bad Delivery Valve
R
A
ECOMMENDED
Replace the Main PCA
Replace the Main PCA
Compressor and Main PCA
Allow the unit to cool off
Real Time temp Driver above 75° C
Battery Temp above 58° C
Allow the unit to cool to below 74° C
and allow the battery to cool below 58°
C
Confirm real time reading for the
temperature.
Restricted air flow
Check fan
Replace the battery if defective.
Allow the unit to cool off below 58° C
Recharge the battery/ defective
charger
Replace the battery
Plug in the unit or defective battery
Turn the unit off first then unplug the
unit from AC Power.
Timers need to reset
Invalid power off
Replace the Main PCA
Check Fan Connections
Replace the fan
Normal event
Straighten the Cannula
Replace internal tubing as needed
Replace the Delivery Valve
Replace the Delivery Valve