Guidance and Manufacturer's Declaration - Electromagnetic Immunity
SimplyGo is intended for use in the electromagnetic environment specified below. The user of SimplyGo
and SimplyFlo should assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Conducted RF
3 Vrms
IEC 61000-5-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a Field strength from fixed transmitters such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
Device is used exceeds the applicable RF compliance level above, the Device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the Device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
c SimplyGo and SimplyFlo incorporate a robust design and conform with Radiated RF Immunity Test up to
10 V/m between 80 MHz and 2.5 GHz while the test requirement is 3 V/m.
Compliance
Electromagnetic Environment - Guidance
Level
Portable and mobile RF communications equipment should be
used no closer to any part of the Device, including cables, that
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
3V
d = 1.2
150 kHz to 80 MHz
P
d = 3.5
c
P
10 V/m
d = 7
800 MHz to 2.5 GHz
P
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with
the following symbol:
80 MHz to 800 MHz
a
should be less than the
1097626, V
. 03