Temperature - Fluke 28 II Ex Getting Started Manual

Hide thumbs Also See for 28 II Ex:
Table of Contents

Advertisement

Temperature

Range
-200 °C to +1090 °C
-328 °F to +1994 °F
[1]
Does not include error of the thermocouple probe.
[2]
Accuracy specification assumes ambient temperature stable to ± 1 °C. For ambient temperature changes of ± 5 °C, rated accuracy
applies after 2 hours.
AC Current
Function
Range
600.0 μA
μA ac
6000 μA
60.00 mA
mA ac
[2]
400.0 mA
6.000 A
A ac
[3,4]
10.00 A
[1]
AC conversions are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range. (5 % to 100 %
of range) and 10 A range (15 % to 100 % or range).
[2]
400 mA continuous. 600 mA for 18 hr maximum.
W 10 A continuous up to 35 °C. <20 minutes on, 5 minutes off at 35 °C to 55 °C. >10 A to 20 A for 30 seconds maximum, 5 minutes
[3]
off.
[4]
>10 A accuracy unspecified.
Resolution
0.1 °C
0.1 °F
Resolution
Burden Voltage
0.1 μA
100 μV/μA
1 μA
100 μV/μA
0.01 mA
1.8 mV/mA
0.1 mA
1.8 mV/mA
0.001 A
0.03 V/A
0.01 A
0.03 V/A
True-rms Digital Multimeter
Detailed Specifications
[1,2]
Accuracy
±(1.0 % + 10)
±(1.0 % + 18)
Accuracy
[1]
(45 Hz – 2 kHz)
±(1.0 % + 2)
27