Page 1
Isothermal Technology Limited, Pine Grove, Southport, PR9 9AG, England Tel: +44 (0)1704 543830 Fax: +44 (0)1704 544799 Internet: www.isotech.co.uk E-mail: info@isotech.co.uk The company is always willing to give technical advice and assistance where appropriate. Equally, because of the programme of continual development and improvement we reserve the right to amend or alter characteristics and design without prior notice.
Page 2
SERVICING THE 3 ZONE METROLOGICAL FURNACE ................. 37 THERMOMETRIC FIXED POINTS - A TUTORIAL ..................38 DEFINING FIXED POINTS AND RELATED DATA ..................40 GENERAL NOTE ON ISOTECH METAL FREEZE POINT CELLS ..............41 PRECAUTIONS TO PREVENT DEVITRIFICATION OF QUARTZ ENVELOPES ..........42 GENERAL COMMENT ..........................43 ADDITIONAL SERVICES AND INFORMATION .....................
INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will be prolonged if regular maintenance and cleaning to remove general dust and debris is carried out. ISOTHERMAL TECHNOLOGY LTD.
93/68/EEC. To ensure emission compliance please ensure that any serial communications connecting leads are fully screened. The product meets the susceptibility requirements of EN 50082-1, criterion B. Symbol Identification Publication Description ISO3864 Caution (refer to manual) IEC 417 Caution, Hot Surface Page 4 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
The Blue conductor should be connected to Neutral and the Brown conductor to Live (Line). Warning: Internal mains voltage hazard. Do not remove the panels. There are no user serviceable parts inside. Contact your nearest Isotech agent for repair. Voltage transients on the supply must not exceed 2.5kV.
If cased, do not return the apparatus to its carrying case until the unit has cooled. There are no user serviceable parts inside. Contact your nearest Isotech agent for repair. Ensure materials, especially flammable materials are kept away from hot parts of the apparatus, to prevent fire risk.
TO REPLACE THE MAIN ELECTRICAL FUSE OR TO REPLACE THE NEUTRAL LINK IT IS NECESSARY TO REMOVE THE REAR PANEL, SEE WARNING ABOVE. THE PANEL WILL BECOME FREE AFTER THE FOUR CORNER SCREWS ARE REMOVED. TWO SPARE FUSES ARE SUPPLIED WITH THE FURNACE. Page 7 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Three zone furnaces can offer wider operating ranges and still meet the requirements for “Optimal Realization of the Defining Points of the ITS-90...” CCT/2000-13. In place of a heat pipe the Model 465 3 Zone High Temperature Furnace uses top and bottom zone heaters to minimize temperature gradients.
An Ethernet interface allows the furnace temperature to be monitored across a network whilst a USB Interface allows programs to be copied or for the furnace heat up and cool down history to be exported. Page 9 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
The cooling coil is accessed from the rear panel (Figure 3) and this will need to be removed to make any connections or service of the connections. The cooling medium can be any number of substances, but in most cases water will suffice. Isotech would recommend a cooled re-circulation system for environmental reasons, however, a simple connection to a water tap will produce the same results.
On the second press the furnace will make a noise as the main control circuit powers up. The main controller will now boot up, starting with the ISOTECH splashscreen. The furnace is now active and ready for use. On the main controller, press the SCROLL button to highlight the Setpoint box (Figure 8).
5. If the controller is in the alarm condition and the complete furnace is powered down, the controller will “remember” the condition and will still need to be manually reset when powered up again. Page 14 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
8. Select Yes and enter with SCROLL button to display the SUPERVISOR screen (Figure 6) 9. When logged in as the SUPERVISOR the equipment will control in the normal way Page 15 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
The HOME page: Figure 8 Furnace temperature Furnace setpoint Bottom Zone temperature Setpoint adjust box Top Zone temperature Data Record indication PAGE button SCROLL button DOWN button UP button Page 16 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
To change the Setpoint of the equipment: 1. From the ISOTECH homepage, press the SCROLL button once, this will highlight the SETPOINT ADJUST box. 2. Press the SCROLL button once more and this will allow adjustment of the setpoint via the UP/DOWN buttons.
This will need to be regularly checked for errors and can easily be corrected within the controller using the following procedure. Figure 11: illustrates a simplified graphical view of the process. Isotech standard offset values are set at the following temperatures: 1. 230°C 2. 440°C 3.
Page 19
8. Scroll to the ISOTECH tab and enter with the SCROLL button 9. Scroll to CAL STATUS and enter with the SCROLL button 10. Scroll to POINT 1 and enter the value from the reference thermometer The controller is now ready to accept a value. Ensure the equipment is thermally stable, you will now need to use the temperature value displayed by the reference thermometer.
6. The controller will now begin the tuning process During the tuning process, AT can now be seen to the left of the clock on the ISOTECH homepage. To disable the tuning feature during a tune, repeat the process above and select OFF. This will now disable the tune and install the previous values.
1. Log in as the SUPERVISOR 2. Plug a suitable USB memory stick to the socket on the control panel 3. Ensure the controller has accepted the memory stick by the icon in the lower left of the ISOTECH homepage 4.
The programs can either be created using the available software, or using the control interface on the equipment. The software can be downloaded at http://www.isotech.co.uk/downloads. Once created, a program can be sent directly to the instrument via the network connection or saved to a USB memory stick and imported from there via the USB socket on the controller panel.
Page 23
Select ‘Per Second’, ‘Per Minute’ or ‘Per Hour’ for ramp timing units. Ramp Units can be edited only when the program is in Reset mode. Ch2 Ramp Units As for ‘Ch1 Ramp Units’ above. Page 23 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
12. Press SCROLL once to highlight the RESET box 13. Scroll to TEST_1 and press the SCROLL button to initiate the program The page will now change to show the current program running. Page 24 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Holdback can be configured to hold either the segment or the complete program. The USER defined value is used to define the actual value applied to it. Page 25 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Page 26
4. Scroll to CH1 HOLDBACK and enter the desired parameter and enter with the SCROLL button 5. Scroll to CH1 HOLDBACK VAL and select the desired temperature, enter using the SCROLL button Page 26 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Page 27
High-purity graphite sleeve High-purity metal Fused-quartz envelope, filled to give a pressure of 1 standard atmosphere at the metal freezing temperature Sealed Metal Freezing-Point Cell Body Dimensions Figure 16: Page 27 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Page 28
Furnace and Holder Diagram Figure 17: Page 28 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Page 29
Page 29 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Page 30
Page 30 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Th , etc. At the temperatures of Tin, Lead and Zinc it is generally permissible to withdraw a thermometer, of type Isotech 909, directly into room temperature. After exposure to Aluminium and Silver point temperatures, the thermometer must be cooled slowly to 450°C.
After 2 minutes remove the rod and replace the thermometer. When the monitoring thermometer has shown no change for some minutes, the plateau has been achieved. A typical melt/freeze sequence is shown in Figure 18. Page 32 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
There follows later a tutorial about Fixed Points. Page 33 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Page 34
Depth of supercool (particularly pronounced for Tin) Plateau showing constant temperature during freeze Temperature drop at end of freeze plateau down to the set point of the furnace controller Page 34 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
CELL HANDLING In order to facilitate introduction to, and removal from the furnace, Isotech can provide for each cell, supplementary equipment largely comprising an Inconel basket with detachable handle. To prevent the cell-surface becoming discoloured, it is recommended that, before using the cell, the basket and insulation be placed in the furnace and the furnace be taken to above the cell working temperature for at least 2 hours.
Page 38 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
Page 39
1 standard atmosphere pressure, for the various metals (and for mercury and water at their triple points), are given in the table on page 26. For a given column height (of the order of 200mm for Isotech sealed freeze point cells), the correction will be proportional to metal density and to a coefficient expressing the sensitivity to pressure of the phase-change temperature.
Furnace with potassium heat-pipe for zinc freezing point. Furnace with either potassium or sodium heat-pipe at aluminium and silver freezing points. Page 40 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
(for example, 14 SWG Nichrome) may be temporarily attached, for use in the 3 Zone Furnace Model 465, the cell holder is inconel. (For use in the 465 furnace the inconel basket for Aluminium, Silver and Copper has an earth lead attached which must be connected during use to the connector provided on the furnace) see diagram.
PRECAUTIONS TO PREVENT DEVITRIFICATION OF QUARTZ ENVELOPES The crucibles (containing the metal) of Isotech sealed fixed point cells are encased in an envelope of pure fused quartz, whose purpose is to avoid contamination of the enclosed metal, by foreign metal ions or oxygen. To this end, it contains an inert gas whose pressure is 1 standard atmosphere at the metal freezing temperature.
(as a system) to ensure that the plateau is clearly identifiable and sufficiently consistent. Page 43 of 45 3 Zone Metrological Furnace 465 Iss.13 – 11/16...
ADDITIONAL SERVICES AND INFORMATION Isotech operates one of the world’s most comprehensive UKAS supervised Laboratories. Training is available to customers at an agreed daily rate. The Isothermal Journal of Thermometry is a must for all Laboratory Managers and is available for purchase.
This application allows logging of data from the ADVANCED model. Additionally if a license is purchased it allows for fully automatic calibration including certificate printing and the calculation of coefficients. Refer to the Isotech I-Cal EASY handbook and help system for more details PROGRAM EDITOR...