Quick Reference Guide
Diagnostic Display
ON/OFF
Illuminates when the Main Power
On/Off Power switch is in the "ON"
(I) position.
PURGE BUTTON
Illuminates when Purge function
is activated. To activate Purge,
remove air polishing insert from
handpiece, turn the Handpiece
Lavage Control to maximum water
flow, press the Purge Button on
the Diagnostic Display. Water will
purge through system lines for two
minutes. To deactivate during two
minute cycle, press Purge button
again or press Tap-On
Pedal.
SERVICE
Illuminates when the system is not
functioning properly. This display
has three distinct modes:
• Slow blink (1 blink per
second) means the system is
not operating within factory
specifications.
• Fast blink (3 blinks per second)
indicates an improper set-up.
• Steady light indicates the system
is overheating.
LOW BATTERY
Illuminates when the Tap-On
Foot Pedal battery power is
approaching end of life. Replace
batteries as instructed in
section 7.9.
Foot
TM
TM
4
Prophy Cycles
Prophy Mode
MANUAL*
Prophy Mode Auto
Cycles
SHORT
MEDIUM
LONG
**
*The 'MANUAL' selection does not automatically cycle between air polish
and rinse, it allows user to air polish conventionally. Press foot pedal to
first position for rinse and then press foot pedal to second position for
air polishing. Continue to repeat as needed.
**'SHORT', 'MEDIUM', and 'LONG' selections will automatically cycle
between air polish and rinse at approximately the times listed in the
table above.
Air Polish
Rinse
Press foot
Press foot
pedal to first
pedal to second
position
position
Approximate Cycle Times
Air Polish
Rinse
0.75 sec
**
1.25 sec
**
2.0 sec
1.0 sec
3.0 sec
2.0 sec