Technical Alarm Messages (INOPs)
INOP Message, Indication
tcpO
(or tcpCO
2
UNPLUGGED
Numeric is replaced by -?-
INOP tone.
TELE CONFIG UNSUPP
INOP tone
!!/!!!TELE DISCONNECT
INOP tone
TELE EQUIP MALF
INOP tone
TELE INCOMPATIBLE
!!/!!! TELE INOP
Severe yellow/red INOP tone.
Tele Sync Unsupp.
INOP tone
TELE UNSUPPORTED
INOP tone
Δ
<
Temp> CHK SOURCES
Numeric is replaced by -?-
Δ
<
Temp> CHK UNITS
Numeric is replaced by -?-
<Temp label> DEACTIVATED
INOP tone
<Temp label> EQUIP MALF
Numeric is replaced by -?-
INOP tone.
<Temp label> NO
TRANSDUCER
Numeric is replaced by -?-
INOP tone.
Source
or tcGas)
tcGas
2
Monitor
Telemetry
Monitor
Monitor
Telemetry
Telemetry
Monitor
TEMP
Difference
TEMP
Difference
TEMP
TEMP
TEMP
5 Patient Alarms and INOPs
What to do
The measurement is switched on but the module is unplugged.
The measurement automatically disappears from the display.
Silencing this INOP switches off the measurement.
Telemetry device not supported (companion mode)
Telemetry transceiver was disconnected or short range radio
link was lost.
For cable connections; check Telemetry interface, cable
connection and setup.
For short range radio connections: if the telemetry transceiver
has not moved out-of-range, check for interference sources
close to the monitor (bluetooth devices, DECT phones,
cellular phones, microwaves, etc.) If this INOP persists, ask
your service personnel to survey the interference sources.
The telemetry device has a malfunction. Disconnect and
reconnect the telemetry device. If the INOP reappears, replace
the telemetry device.
SRR-enabled telemetry device is not supported by this central
software revision. Please check configuration.
Check for further details at the Information Center or in the
Telemetry Data window on the monitor.
The MMS in use does not support synchronization of ECG
and SpO
settings between the monitor and central station
2
after a telemetry device has been paired. Use an MMS with
revision E.0 or above.
This telemetry device is not supported for direct connection to
the monitor.
Not all measurements or values required to perform the
calculation are available. Check measurement sources.
The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
A Temp measurement label in the measurement device has
been deactivated, either by connecting a Pressure transducer in
the shared Press/Temp socket, or by deactivating the label in
the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either reconnect a Temp
transducer or reactivate the measurement label in the
Measurement Selection window.
Contact your service personnel.
The temperature hardware is faulty.
Make sure the TEMP probe is connected to the MMS or
module.
If you silence this INOP, the measurement will be switched off.
81