5 Patient Alarms and INOPs
INOP Message, Indication
SVR/SVRI SET CVP USED
Numeric is replaced by -?-
T/T1/T2/T3/T4 INOPs
TAAP DISABLED
INOP tone
Tamb INOPs
Tart INOPs
Tblood NO TRANSDUC
Numeric is replaced by -?-
INOP tone
Tblood OVERRANGE
Numeric is replaced by -?-
Tcereb INOPS
Tcore INOPs
tcpO
(or tcpCO
2
CAL FAILED
Numeric is replaced by -?-
INOP tone.
tcpO
(or tcpCO
2
CAL REQUIRD
Numeric is replaced by -?-
INOP tone.
tcpO
(or tcpCO
2
CAL RUNNING
Numeric displays first -?- , then
numeric is displayed with a ?
tcpO
(or tcpCO
2
CHECK TIME
tcpO
(or tcpCO
2
CHANGE SITE
If Heat Switch Off is configured to Yes,
numeric is replaced by -?-
INOP tone.
tcpO
(or tcpCO
2
EQUIP MALF
Numeric is replaced by -?-
INOP tone.
tcpO
(or tcpCO
2
NO TRANSDUC
Numeric is replaced by -?-
INOP tone.
tcpO
(or tcpCO
2
STABILIZING
Numeric is displayed with a ?
80
Source
SVR/SVRI
TEMP
Monitor
TEMP
TEMP
C.O.
C.O.
TEMP
TEMP
or tcGas)
tcGas
2
or tcGas)
tcGas
2
or tcGas)
tcGas
2
or tcGas)
tcGas
2
or tcGas)
tcGas
2
or tcGas)
tcGas
2
or tcGas)
tcGas
2
or tcGas)
tcGas
2
Technical Alarm Messages (INOPs)
What to do
A CVP value is required for this calculation, but is not
currently being measured. The monitor is using the CVP value
preset in the
menu.
Setup SVR
See <Temp label> INOPs (under Temp)
The currently selected telemetry configuration on the monitor
does not allow connection of telemetry devices to the monitor.
See <Temp label> INOPs (under Temp)
See <Temp label> INOPs (under Temp)
No transducer attached to the module or catheter
disconnected.
Tblood out of range 17°C - 43°C.
See <Temp label> INOPs (under Temp)
See <Temp label> INOPs (under Temp).
A calibration failed. Check the calibration unit, gas pressure,
and tubing connections, then restart the calibration. If the
calibration has failed more than once, remembrane the
transducer and restart the calibration. If this INOP persists,
contact your service personnel.
Calibration is required before applying the transducer to the
patient.
Insert a membraned transducer into the calibration chamber on
the module, connect the calibration unit to the calibration
chamber, open the gas valve and start the calibration. If this
INOP occurs during a calibration, there may be a module or
transducer malfunction: contact your service personnel.
Wait until the tcpO
/tcpCO
2
2
Site Timer due to time out in 15 minutes or less.
Site Timer has timed out. Change the application site to avoid
skin burns. To reset the Site Timer, either calibrate and change
the measurement site, or change the measurement site and reset
the Site Timer manually by selecting the appropriate site time
from the
menu.
Setup TCGas
There is a malfunction in the transducer or module. Connect
another transducer. If this INOP persists, contact your service
personnel.
No transducer is connected to the tcpO
Silencing the alarm switches off the measurement.
The transducer has not yet reached the selected temperature
and/or skin hyperemization is not yet finished. This INOP will
disappear within three minutes.
calibration is finished.
/tcpCO
module.
2
2