5 Patient Alarms and INOPs
INOP Message, Indication
RA LEAD OFF
Numeric is replaced by -?-
INOP tone.
RAP INOPS
Rem.Alarmdev.Malf
INOP tone
Rem. Display Malf.
!!/!!! REPLACE BATTERY T
Severe yellow/red INOP tone.
During this INOP, alarms cannot be
paused or switched off.
RESP EQUIP MALF
Numeric is replaced by -?-
INOP tone.
RESP ERRATIC
Numeric is replaced by -?-
RESP LEADS OFF
Numeric is replaced by -?-
INOP tone.
RL LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
Settings Malfunc.
INOP tone.
<SO
Label> CAL FAILED
2
Numeric is replaced by -?-
INOP tone.
<SO
Label> CAL MODE
2
Numeric is replaced by -?-
INOP tone.
<SO
Label>CAL REQUIRED
2
Numeric is replaced by -?-
INOP tone.
<SO
Label> CANNOT MEAS
2
Numeric is replaced by -?-
INOP tone.
<SO
Label>CONFIG ERROR
2
Numeric is replaced by -?-
INOP tone.
<SO
Label> CONN OPTMOD
2
Numeric is replaced by -?-
INOP tone.
76
Source
What to do
ECG
The RA electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
Lead Setup
new lead set.
PRESS
See <Pressure label> INOPS (under Pressure).
Monitor
There is a problem with the connection to the remote alert
device. Contact your service personnel to check the remote
alert device and its connections.
Remote
There is a problem with an input device at the remote display.
display
Perform a visual and functional check of all input devices.
Contact your service personnel.
Telemetry
The battery in the telemetry device is almost empty and must
be replaced.
RESP
Contact your service personnel. The RESP hardware is faulty.
RESP
The monitor has detected too many artifacts in the measured
Resp signal. Check that the RA and LL electrodes are correctly
attached and have not dried out.
RESP
Not all the required leads for Resp monitoring are attached.
Make sure that the RA and LL leads are attached.
ECG
The RL electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
Lead Setup
new lead set.
Monitor
The monitor cannot use the predefined settings for
monitoring. Contact your service personnel.
SO
The calibration failed. Check the catheter-to-Optical-Module
2
connection. Manually restart the calibration. Try another
catheter and Optical Module. If the catheter is already inserted,
perform an in-vivo calibration.
SO
Pre-insertion calibration is complete, but the catheter tip is still
2
inside the optical reference. The catheter is now ready for
insertion.
SO
There is no valid calibration data in the Optical Module.
2
Perform either a pre-insertion or an in-vivo calibration.
SO
The signal is out of the normal range, and no oxygen saturation
2
can be derived. Perform an in-vivo calibration. If the INOP
persists, try another Optical Module and catheter.
SO
The Optical Module has been configured to SaO
2
Change to Venous in the setup menu to reconfigure to venous
saturation mode.
SO
The Optical Module was disconnected during data storage.
2
Reconnect the Optical Module for at least 20 seconds.
Technical Alarm Messages (INOPs)
in the
menu to confirm the
Setup ECG
in the
menu to confirm the
Setup ECG
New
New
Mode. Use
2