Technical Alarm Messages (INOPs)
INOP Message, Indication
MSL Power High
MSL Power Off
INOP tone.
MSL Power Overload
INOP tone.
NBP Deactivated
INOP tone.
NBP EQUIP MALF
Numeric is replaced by -?-
INOP tone.
NBP INTERRUPTED
Numeric is replaced by -?-
INOP tone.
NBP MEASURE FAILED
Numeric may be displayed with a -?-
INOP tone.
No Central Monit.
INOP tone
!!/!!! NO DEVICE DATA
NO DEVICE DATA
NO ECG AT CENTRAL
NO ECG SOURCE
Source
What to do
Monitor
The power consumption of the devices connected to the
Measurement Link (MSL) cable is too high. If this situation
continues, the MSL will be switched off. Contact your service
personnel.
Monitor
The power consumption of the devices connected to the
Measurement Link (MSL) cable was too high for too long and
the MSL has been switched off. Contact your service personnel.
Monitor
The power consumption of the devices connected to the
Measurement Link (MSL) cable is much too high or there has
been a short circuit. The MSL has been switched off. Contact
your service personnel.
NBP
The NBP measurement label in the measurement device has
been deactivated by deactivating the label in the Measurement
Selection window. The measurement automatically disappears
from the display. To switch the measurement on again,
reactivate the measurement label in the Measurement Selection
window.
NBP
Remove the cuff from the patient. The NBP hardware is faulty.
Contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
SmartKey is selected.
All
NBP
Check the tubing and cuff for leakages or kinks. Check that
you are using the correct cuff size and placement, and that the
correct patient category is selected. Try restarting the
measurement.
If the INOP occurs repeatedly, contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
SmartKey is selected.
All
This INOP arises when the measurement needed longer than
the maximum time for inflation, deflation or the total
measurement.
NBP
Check that you are using the correct cuff size and placement,
and that the correct patient category is selected. Try restarting
the measurement.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
SmartKey is selected.
All
Check the condition and suitability of the patient for NBP
monitoring. Use another cuff to continue measuring.
Monitor
There is a problem with the communication to the network.
Central monitoring is currently not possible (no patient alarms
or information). Check the connection. Contact your service
personnel.
IntelliBridge Communication with connected device has been lost.
Monitor
The ECG measured with the monitor ECG is not being sent to
the Information Center via the telemetry device.
Telemetry
A telemetry device is paired with the monitor but the
Information Center is not detecting a valid ECG signal from
either of them.
5 Patient Alarms and INOPs
Stop
Stop
Stop
73