5 Patient Alarms and INOPs
INOP Message, Indication
CO
CHANGE SCALE
2
CO
CHK ADAPTER
2
Numeric is replaced by a -?-
INOP tone.
CO
CHECK CAL
2
Numeric is replaced by a -?-
INOP tone.
CO
DEACTIVATED
2
INOP tone.
C.O. DEACTIVATED
INOP tone.
C.O. EQUIP MALF
Numeric is replaced by a -?-
INOP tone.
CO
EQUIP MALF
2
Numeric is replaced by -?-
INOP tone.
CO
FAILED CAL
2
Numeric is replaced by -?-
INOP tone.
CO
NO SENSOR
2
Numeric is replaced by -?-
INOP tone.
CO
NO TRANSDUC
2
Numeric is replaced by -?-
INOP tone.
CO
NO TUBING
2
Numeric is replaced by -?-
INOP tone.
(!!/!!!)CO
Numeric is replaced by a -?-
INOP tone.
CO
OVERRANGE
2
Numeric is replaced by -?-
INOP tone.
CO
PUMP OFF
2
Numeric is replaced by a -?-.
68
Source
CO
2
CO
2
CO
2
CO
2
C.O.
C.O.
CO
2
CO
2
CO
2
CO
2
CO
2
CO
OCCLUSION
2
2
CO
2
CO
2
Technical Alarm Messages (INOPs)
What to do
The CO
wave is clipped. Select a more appropriate wave scale
2
to display the whole wave.
Check that the sensor is connected to the airway adapter, clean
the airway adapter, if necessary. Perform a zero calibration. If
the INOP persists, contact your service personnel.
The CO
value is outside the measurement range. Perform an
2
accuracy check for both calstick cells and, if necessary,
recalibrate the transducer.
The CO
measurement label in the measurement device has
2
been deactivated by deactivating the label in the Measurement
Selection window. The measurement automatically disappears
from the display. To switch the measurement on again,
reactivate the measurement label in the Measurement Selection
window.
The Cardiac Output measurement label in the measurement
device has been deactivated by deactivating the label in the
Measurement Selection window. The measurement
automatically disappears from the display. To switch the
measurement on again, reactivate the measurement label in the
Measurement Selection window.
There is a problem with the C.O. hardware. Contact your
service personnel.
The Measurement Extension is faulty. Unplug and replug the
Multi-Measurement Module with Extension. If you are using
the mainstream method, unplug and replug the transducer or
try another transducer. If the INOP persists, contact your
service personnel.
Make sure that the Cal cell was changed between CAL1 and
CAL2. Repeat the calibration. If the INOP reappears, try
another transducer. If the INOP persists, contact your service
personnel.
There is no CO
sensor connected. If you silence this INOP
2
the CO
measurement will be switched off.
2
There is no CO
transducer connected. If you replace the
2
transducer, the new transducer must be calibrated. If you
silence this INOP the CO
measurement will be switched off.
2
Either the sample line is disconnected, or an incorrect line is
attached. Check the connection. If necessary, connect another
sample line (Use only the approved accessories).
If you silence this INOP, the measurement will be switched off.
The sample line or exhaust tube is blocked. Check the tubing,
then disconnect and reconnect the sample line. If the INOP
persists, connect a new sample line.
The CO
value is higher than the measurement range. If you
2
suspect a false high value, contact your service personnel.
The pump has been switched off for fifteen minutes. To switch
it on again, select Pump On in the Setup CO
menu.
2