5 Patient Alarms and INOPs
INOP Message, Indication
CENTRAL:TELE ONLY
INOP tone
Charge BATT1/BATT2 now
INOP tone
CHARGER MALFUNC
INOP tone, battery LED may flash
Check Alarm Lamps
INOP tone.
CHECK BATT TEMP
INOP tone
Check DrugSettings
INOP tone
Check ECG Settings
INOP tone
!!CHECK ECG SOURCE
INOP tone
Check Flex Texts
INOP tone
CheckInternVoltage
at Information Center
Check Keyboard
INOP tone
Check Main Board 2
INOP tone.
Check Monitor Func
INOP tone.
Check Monitor Temp
INOP tone
Check Mouse Device
INOP tone.
Check MSL Voltage
INOP tone
66
Source
What to do
Monitor
System connectivity via telemetry device is limited (No alarms,
only local numerics) when in companion mode and host
monitor does not have system connectivity. Only telemetry
device parameters can be displayed at central station.
Batteries
Battery must be charged. Connect the monitor to mains power
or exchange the battery.
Batteries
There is a problem with the battery charger in the monitor.
Connect the monitor to mains power and contact your service
personnel.
Monitor
Perform a visual check of the alarm lamp to establish whether
there is a problem. Contact your service personnel to check the
internal connections to the alarm lamps.
Battery
The temperature of one or both batteries is too high. Check
that ventilation openings (if applicable) are not blocked and
monitor is not exposed to heat.
Monitor
There was a problem loading the drug settings. Check that the
settings are complete and correct.
Telemetry
Synchronization of ECG settings between the monitor and
Information Center has failed. Check that the ECG settings in
use are appropriate.
Monitor
The telemetry device and the monitor both have valid ECG
signals. Unpair the telemetry device and the monitor if they are
no longer used for the same patient.
Monitor
Check the names of the monitor menus, for example the labels
for screens, profiles, event or trend group names, before you
resume monitoring. If they are unexpected, there may be a
problem with the monitor software. Contact your service
personnel.
Monitor
Potential problem with alarm lamps, display or interfaces
detected. Contact your service personnel. This INOP will
appear on the monitor as
Monitor
Perform a visual and functional check of the keyboard. Contact
your service personnel.
Monitor
There is a problem with the second main board in the monitor.
Contact your service personnel.
Monitor
Potential problem with alarm lamps, display or interfaces
detected. Contact your service personnel. This INOP may
appear on the Information Center as
CheckInternVoltage
Monitor
The temperature inside the monitor is too high. Check that the
monitor ventilation is not obstructed. If the situation
continues, contact your service personnel.
Monitor
Perform a visual and functional check of the mouse input
device. Contact your service personnel.
Monitor/
There is a problem with the voltage of the Measurement Link
Multi-
(MSL). Contact your service personnel.
Measuremt
Module
Technical Alarm Messages (INOPs)