Download  Print this page

Advertisement

Table of Contents
Technical Alarm Messages (INOPs)
INOP Message, Indication
CCI NO BSA
CCI numeric unavailable
INOP tone
CCO BAD PRESS SIGN
numeric is replaced by -?-
INOP tone
CCO NO <Pressure label>
numeric is replaced by -?-
INOP tone may sound
CCO NO CALIBRATION
numeric is replaced by -?-
CCO NO PRESS
at Information Center
CCO NOT SUPPORTED
numeric is replaced by -?-
INOP tone
CCO/CCI OVERRANGE
numeric is replaced by -?-
INOP tone
CCO <Pressure label>
INVALID
numeric is replaced by -?-
INOP tone may sound
CCO PRESS INVALID
at Information Center
CCO PRESS OVERRANG
numeric is replaced by -?-
INOP tone
CCO PULSE OVERRANG
numeric is replaced by -?-
INOP tone
CCO/Tbl NO TRANSD
Numeric is replaced by -?-
INOP tone
CCO RECALIBRATE
numeric is replaced by -?-
Source
What to do
C.O.
CCI cannot be calculated because the patient's body surface
area is unknown. Enter the patient weight and height to
provide the BSA for CCI calculation.
C.O.
The arterial pressure wave can currently not be used for pulse
contour calculation for CCO or CCI measurement. Possible
causes are air bubbles in the tubing or a physiological
condition, for example severe arrhythmia.
C.O.
CCO/CCI cannot be calculated. Make sure that the pressure
chosen in the
Setup CCO
matches the pressure measured with the arterial catheter for
CCO measurement. A pressure from an external device cannot
be used. Select another pressure label, either ABP, Ao, ART,
BAP, FAP, or UAP.
C.O.
The CCO measurement is currently not calibrated.
C.O.
CCO/CCI cannot be calculated. Make sure that the pressure
chosen in the
Setup CCO
matches the pressure measured with the arterial catheter for
CCO measurement. A pressure from an external device cannot
be used. Select another pressure label, either ABP, Ao, ART,
BAP, FAP, or UAP.
C.O.
A catheter for transpulmonary C.O. measurements has been
unplugged and replaced with a Right Heart C.O. catheter, or
the measurement mode has been changed manually. Silencing
this INOP switches the measurement off.
C.O.
The measured CCO or CCI value is not within the specified
range for CCO/CCI measurement.
C.O.
The arterial pressure selected for pulse contour calculation for
CCO is available but currently invalid. Make sure the pressure
transducer is connected and the zero calibration is valid.
C.O.
The arterial pressure selected for pulse contour calculation for
CCO is available but currently invalid. Make sure the pressure
transducer is connected and the zero calibration is valid.
C.O.
The mean value of the arterial pressure values used for pulse
contour calculation for CCO is below 0 mmHg or above 300
mmHg.
C.O.
The pulse rate of the pressure used for pulse contour
calculation for CCO is below 30 bpm or above 240 bpm.
C.O.
No transducer attached to the module or catheter
disconnected.
C.O.
The most recent CCO or CCI calibration was made over 8
hours ago or the arterial pressure measurement used for CCO
calculation has been zeroed after the CCO calibration was
performed. You should recalibrate CCO or CCI with
transpulmonary C.O. measurements at least every 8 hours or
when the hemodynamic condition of the patient has changed.
The pressure measurement must be zeroed before a CCO
calibration.
5 Patient Alarms and INOPs
menu under
CCO From
menu under
CCO From
65

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0
Latest comments: