Safety and Performance Tests
Recommended Separation Distance
The monitor, equipped with a wireless network interface, intentionally receives RF electromagnetic
WARNING
energy for the purpose of its operation. Therefore, other equipment may cause interference, even if that
other equipment complies with CISPR emission requirements.
In the following table, P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation distance in metres
(m). The values given in brackets are for respiration.
Portable and mobile RF communications equipment should be used no closer to any part of the
monitor, including cables, than the recommended separation distance calculated from the equation
appropriate for the frequency of the transmitter.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with this symbol:
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
1.If ECG/SpO
the range 2.0 to 2.3 GHz where it is 1 V/m.
IEC 60601-1-2
test level
Compliance level
3 V
3 V
RMS
RMS
150 kHz to
(1 V
RMS
80 MHz
respiration)
3 V/m
3 V/m
80 MHz to
(1 V/m for respiration)
2.5 GHz
For short range radio,
see note
signals are acquired from a telemetry device via short range radio the compliance level is 3V/m except in
2
31 Installation and Specifications
Electromagnetic environment guidance
Recommended separation distance:
for
d
=
1,2 P
for respiration:
d
=
3,5 P
Recommended separation distance:
80 MHz to 800 MHz
d
=
1,2 P
1
.
80 MHz to 800 MHz for respiration
d
=
3,5 P
800 MHz to 2,5 GHz
d
=
2,3 P
800 MHz to 2,5 GHz for respiration
d
=
7,0 P
2.0 to 2,3 GHz for short range radio
d
=
7,0 P
:
:
: