31 Installation and Specifications
CO
Alarm
2
Specifications
imCO
High
2
awRR High
awRR Low
Apnea delay
Interfering Gas and Vapor Effects On CO
The specified deviations are valid when the appropriate corrections are switched on and set correctly.
Gas or Vapor
Nitrous Oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Metered dose inhaler
propellants
Desflurane
Ethanol
Isopropanol
Acetone
Methane
320
Range
2 to 20 mmHg
(0.3 to 3.0 kPa)
Adult/pedi: 10 to 100 rpm
Neo: 30 to 150 rpm
Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
10 to 40 seconds
Measurement Values
2
Gas Level
(% volume fraction)
60
4
5
5
5
80
50
-
15
0.1
0.1
0.1
1.0
Adjustment
Delay
steps of 1 mmHg
less than 14 seconds
(0.1 kPa)
under 20 rpm: 1 rpm
less than 14 seconds
steps
over 20 rpm:5 rpm
settings <20 rpm: less than 4 seconds
steps
settings >20 rpm: less than 14 seconds
5 second steps
set apnea delay time +4 seconds
Additional deviation due to gas interference, measured
at 0 - 40 mmHg CO
2
±1 mmHg
±2 mmHg
±2 mmHg
±2 mmHg
±2 mmHg
-5 mmHg
±1 mmHg
not specified for use
+5 mmHg
±1 mmHg
±1 mmHg
±1 mmHg
±1 mmHg
Measurement Specifications