Measurement Specifications
Invasive Pressure Alarm Specifications Range
Tachycardia
Bradycardia
Temp
Complies with EN 12470-4:2000. Specified without transducer.
Temp Performance Specifications
Temp
Average Time Constant
Temp Alarm Specifications
Temp High/Low Alarms
Predictive Temperature
Performance Specifications
Technology
Probe Types
Temperature Measurement Range
Resolution
Accuracy
(Oral, axillary and rectal measurements)
CO
2
The CO
2
system complies with IEC 60601-2-49:2001).
Range
Resolution
Accuracy
Range
–1 to 45°C (30 to 113°F)
measurement complies with EN ISO 21647:2004 + Cor.1:2005 (except alarm system; alarm
Difference to high limit 0 to
50 bpm
Clamping at 150 to
300 bpm
Difference to low limit 0 to
50 bpm
Clamping at 25 to 100 bpm 5 bpm steps
–1 to 45°C (30 to 113°F)
0.1°C (0.2°F)
±0.1°C (±0.2°F)
Less than 10 seconds
Adjustment
-1 to 35°C (30 to 95°F), 0.5°C (1.0°F) steps
35 to 45°C (95 to 113°F), 0.1°C (0.2°F) steps
Welch Allyn® SureTemp Plus®
oral/axillary, rectal
26.7 to 43.3°C (80 to 110°F)
±0.1°C (±0.2°F)
±0.1°C (±0.2°F) (in continuous mode, complies with ASTM 1112-
00
31 Installation and Specifications
Adjustment
5 bpm steps
5 bpm steps
5 bpm steps
Delay
max.
14 seconds
max.
14 seconds
317