Download  Print this page

Advertisement

Table of Contents
Monitor Performance Specifications
Monitor Interface Specifications
Basic Nurse Call
Relay
Wireless Network
Device Interface
(integrated, for
compatible network
options see below)
Short Range Radio
1
Interface
1.The short range radio interface is compatible with the following telemetry devices: TRx4841A/TRx4851A IntelliVue
Telemetry System Transceiver.
Display Specifications
All displays
Integrated SVGA
Display
Video Interface SVGA
Compatible Devices
Displays (must be approved
for medical use)
IntelliVue 802.11 Bedside Adapter (Wireless Network Adapter)
Internal Wireless Adapter
IntelliVue Instrument Telemetry Wireless Network (USA only)
Internal WMTS Adapter
IntelliVue Instrument Telemetry Wireless Network (except USA)
Internal ISM Adapter
Connector
3.5 mm phone jack, active closed contact only
Contact
<=100 mA, <=24 V DC
Isolation
1.5 kV
Delay
<(Configured Latency + 0.5 sec)
Signals
RD+/-, TD+/-: IEEE 802.3 10 Base-T, PWR, GND
12.5 V ±20 %, 3.5 W continuous
Type
Internal SRR interface
Technology
IEEE 802.15.4
Frequency Band
2.4 GHz ISM (2.400 - 2.483 GHz)
Modulation Technique DSSS (O -QPSK)
Effective radiated
max. 0 dBm (1 mW)
power
Sweep Speeds
6.25, 12.5, 25 and 50 mm/s;
Resolution
800 x 600
Refresh frequency
60 Hz
Useful screen
170.4 x 127.8 mm
Pixel size
0.213 x 0.213 mm
Horizontal Frequency
37.5 kHz
Refresh frequency
60 Hz
Video Signals
0.7 Vpp @ 75 Ohm, HSYNC/VSYNC Signals TTL
Connector
15 pin D-SUB
M8031B
M8033C
Technology
Frequency Band
Technology
Frequency Band
Technology
Frequency Band
31 Installation and Specifications
XGA color 15" LCD touchscreen
SXGA color 17" LCD Touchscreen
IEEE 802.11a/b/g
2.4 GHz and 5 GHz ISM Band
compatible with Philips Cellular Telemetry System
(CTS), cellular infrastructure
WMTS, 1395-1400 MHz and 1427-1432 MHz
compatible with Philips Cellular Telemetry System
(CTS), cellular infrastructure
2.4 GHz ISM
307

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0
Latest comments: