Download  Print this page

Advertisement

Table of Contents
Out-Of-Hospital Transport - Standards Compliance
• Specification for degrees of protection provided by enclosures according to IEC/EN 60529: IP 32
• EN 1789 +A1:2003 Medical vehicles and their equipment - Road ambulances (chapter 6 - Medical
Devices).
• Radiated susceptibility 20 V/m according to EN ISO 9919 (SpO
• Altitude Range from -500 to 3000 m operating and -500 to 4600 m storage and transportation.
• Extended radiated susceptibility tests
The MP5 patient monitor with its out-of-hospital parameter set provides a general immunity level
of 20 V/m with only few restrictions. Details are as listed below:
– GSM 900: Immunity at 900 MHz (uplink mobile phone), 20 V/m (ECG:10 V/m), duty cycle
1:8
– GSM 1800: Immunity at 1800 MHz (uplink mobile phone), 20 V/m, duty cycle 1:8.
– DECT: Immunity at 1800 MHz (digital cordless phone), 20 V/m, duty cycle 1:24
– AM: 1 kHz Immunity from 80 MHz to 1.0 GHz (any radio communication unit, broadcasting
and TV transmitter), 20 V/m, modulation factor 80 %. (ECG: 20 V/m except 600-950 MHz
where it is 10 V/m and Temperature which holds 3 V/m over the full range).
Temperature measurement accuracy may be compromised in the presence of strong electromagnetic
CAUTION
fields (>3V/m) in certain small frequency bands.
• Magnetic Field emission according to MIL STD 461E, Chapter RE101: Radiated emissions,
magnetic field, 30 Hz to 100 kHz. Limit class: Army.
• Magnetic Field susceptibility: Radiated susceptibility, magnetic field, 50, 60 and 400 Hz,
18 µT(15 A/m)
• Operating ambient temperature testing over the range from 0 to 40°C (32 to 100°F).
• Operating ambient humidity testing up to 95 % RH at 40°C (100°F), non condensing.
There may be additional requirements for transport situations in air, on water or in difficult terrain in
N O T E
certain countries, e.g. EU.
Physical Specifications
Specification
Maximum Weight
4.0 kg ±5 %
(8.8 lb)
W x H x D
<259 x 248 x 186 mm
10.2 x 9.8 x 7.3 in
<312 x 248 x 186 mm
12.3 x 9.8 x 7.3 in
Comments
with ECG/Resp, NBP, SpO
and battery
2
without Predictive Temperature unit
with Predictive Temperature unit
31 Installation and Specifications
) and EN ISO 21647 (CO
2
).
2
303