Zeroing The Pressure Transducer - Philips IntelliVue MP5 Instructions For Use Manual

Patient monitor.
Hide thumbs

Advertisement

Table of Contents
14 Monitoring Invasive Pressure
2
Select the appropriate label from the list.
Label
ABP
ART
Ao
CVP
ICP
LAP
Extended Pressure Label Set
The following additional labels are available if Label Set is set to Full. This setting can only be
changed in Configuration Mode.
Note that if your monitor is connected to an Information Center, the additional labels in the extended
label set may not be correctly displayed. See the Configuration Guide for your monitor for more
information.
Label
BAP
FAP
IC1, IC2
P1, P2, P3, P4 Alternative non-specific pressure labels

Zeroing the Pressure Transducer

To avoid inaccurate pressure readings, the monitor requires a valid zero. Zero the transducer in
accordance with your hospital policy (at least once per day). You must perform a zero:
• when you use a new transducer or tubing
• every time you reconnect the transducer cable to the monitor
• if you think the monitor's pressure readings are not correct.
Zeroing ICP (or IC1/IC2)
Your hospital guidelines may require you to zero the ICP transducer less frequently than other
transducers, due to the need for aseptic conditions. When you zero an ICP transducer, the zero values
are automatically stored and you will not be prompted to repeat the zero procedure.
If you select the label ICP (or IC1/IC2), the measurement device uses the most recently stored zero.
WARNING
Therefore, make sure you zeroed the transducer correctly in accordance with the transducer
manufacturer's instructions and your hospital policy. When you use a transducer that you cannot
rezero after placement, ensure that you keep the measuring device with the patient so that you are
certain you have the correct zero data for this patient.
176
Description
Arterial blood pressure
Arterial blood pressure (alternative)
Aortic pressure
Central venous pressure
Intracranial pressure
Left atrial pressure
Description
Brachial arterial pressure
Femoral arterial pressure
Alternative intracranial pressures
Zeroing the Pressure Transducer
Label
Description
Non-specific pressure label
P
Pulmonary artery pressure
PAP
Right atrial pressure
RAP
Umbilical arterial pressure
UAP
Umbilical venous pressure
UVP