7 ECG, Arrhythmia, ST and QT Monitoring
You will see that
• if both the V-Tach Heart Rate Limit and the V-Tach Run Limit are exceeded, a red V-Tach alarm is
generated
• if the ventricular heart rate exceeds the V-Tach Heart Rate Limit but not the V-Tach Run Limit, a
yellow Non-Sustain VT alarm is generated.
126
** Non-Sustain VT
**Pair
PVC Run < 8
PVCs
HR > 100
PVC =
2
100
** Run PVCs
PVC Run > 2 but
HR
1
2
3
4
5
Number of Consecutive PVCs (PVC Run)
V-Tach Run Limit
***V-Tach
PVC Run
HR > 100
12
100
6
7
8
9
10
11
Arrhythmia Alarms
8
V-Tach Heart
Rate Limit
** Vent Rhythm
PVC Run > 12
HR
100
12
14
15
Ventricular Rhythm
Run Limit