7 ECG, Arrhythmia, ST and QT Monitoring

ECG Lead Fallback

If fallback is configured on and there is a leads off INOP in the primary lead (and in the secondary
lead, if you are using multi-lead monitoring) for longer than 10 seconds, and if another lead is
available, this available lead automatically becomes the primary lead. This is known as lead fallback.
When the Leads Off condition is corrected, the leads are automatically switched back.
This setting can only be changed in Configuration Mode.
ECG Lead Placements
The labels and colors of the ECG electrodes differ according to the standards that apply for your
hospital. The electrode placement illustrations in this chapter use the AAMI labels and colors.
Electrode labels
AAMI
RA
LA
LL
RL
V
V1
V2
V3
V4
V5
V6
106
Electrode colors
EASI
IEC
AAMI
I
R
White
S
L
Black
A
F
Red
N
N
Green
E
C
Brown
C1
Brown/Red
C2
Brown/Yellow
C3
Brown/Green
C4
Brown/Blue
C5
Brown/Orange
C6
Brown/Violet
ECG Lead Fallback
IEC
Red
Yellow
Green
Black
White
White/Red
White/Yellow
White/Green
White/Brown
White/Black
White/Violet