⑤ Analog Monitor Signals
There are 3 types of monitoring signals : " Motor Current Monitor (MOTOR_I) " , " Torque Load Monitor
(LOAD) " , and " Rotating Speed Analog Monitor Voltage (SPEED_V) " . Please refer to Fig. 38 for
connections.
16 - 2 External Input / Output Connector B
( 1 ) Details of External Input / Output Connector B ④ Signals
WARNING
・ DO NOT connect any circuit other than SELV ( DC+24V) (Safety Extra Low Voltage) to the
External Input / Output Connector B ④ of the CONTROLLER. This will cause I / O board damage
in the CONTROLLER.
・ Do not supply over voltage or over current into the input / output circuit. Always install a LOAD
(resistor) to the output circuit to eliminate the chance of damage to the CONTROLLER.
CAUTION
Input / Output Connector B ④ DOES NOT use Pin No. 7, No. 8, and No. 15. If pin No. 7, No. 8, and
No. 15 are connected, the CONTROLLER will be damage.
Table. 6
Pin
Code
Function
No.
Emergency
1
EMG-INA
Stop A
Motor Connect
2
MT-CNA
Contact A
Safety Relay
3
SAFE- 1 A
Contact 1 A
The side of CONTROLLER
1KΩ
12
Motor Current Monitor ( MOTOR_I)
1KΩ
24
Torque Load Monitor ( LOAD)
1KΩ
25
Rotating Speed Analog
Monitor Voltage (SPEED_V)
13
DC0V
Fig. 38
Input /
Output
External Power
Source input
for Emergency
Input
Stop Signal or
Emergency Stop
Signal ' OFF (Open) '
Continuity, ' OFF
(Open) ', between
Output
Pin No. 2 and Pin
No. 10 the motor is
connected.
Pin No. 3 and Pin
No. 11 continuity
Output
' ON (Closed) '
Safety Relay is OFF
Description
External Power Source input for Emergency
Stop Signal or Emergency Stop Signal.
N o r m a l O p e r a t i o n ' O N ( C l o s e d ) ' ,
Emergency ' OFF (Open) '.
When using the Emergency Stop Signal, set
. (Refer to P 92 " 18 - 4
parameter
Selection of Emergency Stop Function "
section.)
When there is continuity, OFF, between
Pin No. 2 and Pin No. 10 and the selected
motor is connected, if no continuity is
present, the motor is disconnected or the
motor cord is broken.
When there is continuity between Pin
No. 3 and Pin No. 1 are ' ON (Closed) '
Safety Relay is OFF (System Stopped), no
continuity Safety Relay is ' OFF (Open) '
Normal Operation.