4 ECG and Arrhythmia Monitoring

INOP Messages

INOP messages communicate conditions that prevent the device from monitoring or analyzing the
ECG. INOP messages are displayed just above the HR/Arrhythmia alarm status area. If multiple
INOP conditions exist, the associated INOP messages will cycle through, alternating every 2 seconds.
Table 7 ECG INOPs
Alarm Message
Cannot Analyze ECG
ECG Cable Failure
Leads Off
Pads /Paddles Off
ECG Unplugged
ECG Equip Malfunction A malfunction has occurred in the
Pads/Paddles Cable
Failure
Pads ECG Equip
Malfunction
58
Condition
Cannot reliably monitor the ECG in
Wave Sector 1.
During the Operational Check, a short
has been detected between a lead wire
and ground
An electrode used for Wave Sector 1
may be off or not attached securely.
The multifunction electrode pads or
paddles used as the source for the Wave
Sector 1 may be off or not attached
securely.
The primary ECG is derived from leads
and the ECG cable is not connected.
ECG hardware.
During the Operational Check, a
failure was detected in the pads or
paddles cable during the pads/paddles
ECG test.
A device hardware failure was detected. INOP message, INOP tone
Heart Rate and Arrhythmia Alarms
Indication
INOP message, INOP tone
INOP message, INOP tone
INOP message, INOP tone
INOP message, INOP tone
INOP message, INOP tone
INOP message, INOP tone
INOP message, INOP tone