5 Patient Alarms and INOPs
INOP Message, Indication
Tblood NO TRANSDUC
Numeric is replaced by -?-
INOP tone
Tblood OVERRANGE
Numeric is replaced by -?-
Tcereb INOPS
Tcore INOPs
tcpO
2
CAL FAILED
Numeric is replaced by -?-
INOP tone.
tcpO
2
CAL REQUIRD
Numeric is replaced by -?-
INOP tone.
tcpO
2
CAL RUNNING
Numeric displays first -?- , then
numeric is displayed with a ?
tcpO
2
CHECK TIME
tcpO
2
CHANGE SITE
If Heat Switch Off is configured to Yes,
numeric is replaced by -?-
INOP tone.
tcpO
2
EQUIP MALF
Numeric is replaced by -?-
INOP tone.
tcpO
2
NO TRANSDUC
Numeric is replaced by -?-
INOP tone.
tcpO
2
STABILIZING
Numeric is displayed with a ?
tcpO
2
UNPLUGGED
Numeric is replaced by -?-
INOP tone.
TELE CONFIG UNSUPP
INOP tone
78
or tcGas)
(or tcpCO
2
or tcGas)
(or tcpCO
2
or tcGas)
(or tcpCO
2
or tcGas)
(or tcpCO
2
or tcGas)
(or tcpCO
2
or tcGas)
(or tcpCO
2
or tcGas)
(or tcpCO
2
or tcGas)
(or tcpCO
2
or tcGas)
(or tcpCO
2
Source
What to do
C.O.
No transducer attached to the module or catheter
disconnected.
C.O.
Tblood out of range 17°C - 43°C.
TEMP
See <Temp label> INOPs (under Temp)
TEMP
See <Temp label> INOPs (under Temp).
tcGas
A calibration failed. Check the calibration unit, gas pressure,
and tubing connections, then restart the calibration. If the
calibration has failed more than once, remembrane the
transducer and restart the calibration. If this INOP persists,
contact your service personnel.
tcGas
Calibration is required before applying the transducer to the
patient.
Insert a membraned transducer into the calibration chamber on
the module, connect the calibration unit to the calibration
chamber, open the gas valve and start the calibration. If this
INOP occurs during a calibration, there may be a module or
transducer malfunction: contact your service personnel.
tcGas
Wait until the tcpO
tcGas
Site Timer due to time out in 15 minutes or less.
tcGas
Site Timer has timed out. Change the application site to avoid
skin burns. To reset the Site Timer, either calibrate and change
the measurement site, or change the measurement site and reset
the Site Timer manually by selecting the appropriate site time
from the
Setup TCGas
tcGas
There is a malfunction in the transducer or module. Connect
another transducer. If this INOP persists, contact your service
personnel.
tcGas
No transducer is connected to the tcpO
Silencing the alarm switches off the measurement.
tcGas
The transducer has not yet reached the selected temperature
and/or skin hyperemization is not yet finished. This INOP will
disappear within three minutes.
tcGas
The measurement is switched on but the module is unplugged.
The measurement automatically disappears from the display.
Silencing this INOP switches off the measurement.
Monitor
Telemetry device not supported (companion mode)
Technical Alarm Messages (INOPs)
/tcpCO
calibration is finished.
2
2
menu.
/tcpCO
2
module.
2