Advertisement

5 Patient Alarms and INOPs
INOP Message, Indication
!!/!!! REPLACE BATTERY T
Severe yellow/red INOP tone.
During this INOP, alarms cannot be
paused or switched off.
RESP EQUIP MALF
Numeric is replaced by -?-
INOP tone.
RESP ERRATIC
Numeric is replaced by -?-
RESP LEADS OFF
Numeric is replaced by -?-
INOP tone.
RL LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
Settings Malfunc.
INOP tone.
<SO
Label> CAL FAILED
2
Numeric is replaced by -?-
INOP tone.
<SO
Label> CAL MODE
2
Numeric is replaced by -?-
INOP tone.
<SO
Label>CAL REQUIRED
2
Numeric is replaced by -?-
INOP tone.
<SO
Label> CANNOT MEAS
2
Numeric is replaced by -?-
INOP tone.
<SO
Label>CONFIG ERROR
2
Numeric is replaced by -?-
INOP tone.
<SO
Label> CONN OPTMOD
2
Numeric is replaced by -?-
INOP tone.
<SO
Label> EQUIP MALF
2
Numeric displays -?-.
INOP tone
SO
INCOMPATIBLE
2
INOP tone
<SO
Label> IN-VIVO CAL
2
74
Source
What to do
Telemetry
The battery in the telemetry device is almost empty and must
be replaced.
RESP
Contact your service personnel. The RESP hardware is faulty.
RESP
The monitor has detected too many artifacts in the measured
Resp signal. Check that the RA and LL electrodes are correctly
attached and have not dried out.
RESP
Not all the required leads for Resp monitoring are attached.
Make sure that the RA and LL leads are attached.
ECG
The RL electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
Lead Setup
new lead set.
Monitor
The monitor cannot use the predefined settings for
monitoring. Contact your service personnel.
SO
The calibration failed. Check the catheter-to-Optical-Module
2
connection. Manually restart the calibration. Try another
catheter and Optical Module. If the catheter is already inserted,
perform an in-vivo calibration.
SO
Pre-insertion calibration is complete, but the catheter tip is still
2
inside the optical reference. The catheter is now ready for
insertion.
SO
There is no valid calibration data in the Optical Module.
2
Perform either a pre-insertion or an in-vivo calibration.
SO
The signal is out of the normal range, and no oxygen saturation
2
can be derived. Perform an in-vivo calibration. If the INOP
persists, try another Optical Module and catheter.
The Optical Module has been configured to SaO
SO
2
Change to Venous in the setup menu to reconfigure to venous
saturation mode.
SO
The Optical Module was disconnected during data storage.
2
Reconnect the Optical Module for at least 20 seconds.
SO
The SO
/SvO
2
2
and replug the Optical Module and SO
Exchange the modules. If the INOP persists, contact your
service personnel.
SO
The SO
Module or Optical Module is not supported. Contact
2
2
your service personnel.
SO
The in-vivo calibration is not yet complete. Lab values must be
2
stored to the Optical Module to complete the calibration.
Either continue with the next steps of the current calibration or
recall the previous calibration.
Technical Alarm Messages (INOPs)
in the
menu to confirm the
Setup ECG
Module or Optical Module is faulty. Unplug
2
/SvO
2
New
Mode. Use
2
module.
2