Advertisement

5 Patient Alarms and INOPs
Alarm Message
**BIS LOW
***BRADY/P xxx<yyy
or
***BRADY xxx<yyy
**CCO/CCI HIGH
**CCO/CCI LOW
**CPP HIGH
**CPP LOW
***DESAT or
***DESAT xxx<yyy
**etCO
**etCO
**etO
2
**etO
2
*/**/***
EVENT:<GRP>
*/**/*** EVENT
at Information center
***EXTREME BRADY
***EXTREME TACHY
54
From
BIS
Press, SpO
CCO
CCO
CPP
CPP
SpO
2
CO
, Resp,
HIGH
2
2
AGM
CO
, Resp,
LOW
2
2
AGM
O
, AGM
HIGH
2
O
, AGM
LOW
2
Event
surveillance
Event
surveillance
ECG
ECG
Condition
The Bispectral Index value has dropped
below the low alarm limit.
The heart rate from the Pulse signal has
2
fallen below the bradycardia limit. xxx
denotes the lowest measured value; yyy
is the bradycardia limit.
Continuous Cardiac Output or CC
Index is above the high alarm limit.
Continuous Cardiac Output or CC
Index is below the low alarm limit.
The CPP value has exceeded the high
alarm limit.
The CPP value has fallen below the low
alarm limit.
The SpO
value has fallen below the
2
desaturation alarm limit. xxx denotes
the lowest measured value, and yyy is
the desaturation limit.
The end tidal CO
high alarm limit has
2
been exceeded.
The end tidal CO
value has fallen
2
below the low alarm limit.
The end tidal O
high alarm limit has
2
been exceeded.
The end tidal O
value has fallen below
2
the low alarm limit.
An event has occurred and the event
notification is configured to alarm.
<GRP> is the event group
An event has occurred and the event
notification is configured to alarm.
Check on the monitor for more details
on event group.
The bradycardia limit has been
exceeded.
The tachycardia limit has been
exceeded.
Patient Alarm Messages
Indication
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and alarm
limit is highlighted, red alarm
lamp, alarm tone.
numeric flashes and high alarm
limit is highlighted, yellow
alarm lamp, alarm tone.
numeric flashes and low alarm
limit is highlighted, yellow
alarm lamp, alarm tone.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes, red alarm
lamp, alarm tone.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes, and low limit
is highlighted, yellow alarm
lamp, alarm tone.
event group name flashes,
yellow or red alarm lamp and
alarm tone
(on monitor) event group
name flashes, yellow or red
alarm lamp and alarm tone
numeric flashes and alarm
limit is highlighted, red alarm
lamp, alarm tone.
numeric flashes and alarm
limit is highlighted, red alarm
lamp, alarm tone.