4 Alarms
The Review Alarms window pop-up keys appear when the window is opened. If alarm pause
extension is disabled, the pause pop-up keys are inactive.

Latching Alarms

The alarm latching setting for your monitor defines how the alarm indicators behave when you do not
acknowledge them. When alarms are set to non-latching, their indicators end when the alarm
condition ends. Switching alarm latching on means that visual and/or audible alarm indications are still
displayed or announced by the monitor after the alarm condition ends. The indication lasts until you
acknowledge the alarm.
Viewing the Alarm Latching Settings
To see the alarm latching setting for your monitor
1
In the monitor's Main Setup menu, select Alarms.
Select Alarm Settings, and see the Visual Latching and Audible Latching
2
settings.
This setting can only be changed in Configuration Mode. You should be aware of the settings chosen
for your unit. There are three possible choices each for visual and audible latching, Red, Red and
Yellow, and Off. These choices can be combined to give the following settings:
Visual Latching
Audible latching
R = red alarms, Y = yellow alarms
Alarm Latching Behavior
Red and Yellow Measurement
Alarms
Alarm has not been
acknowledged.
Alarm has been
acknowledged.
All INOPs are non-latching. See "Yellow Arrhythmia Alarms" on page 112 for information on one-star
yellow alarms latching behavior.
50
Alarm
Limits
R&Y
R&Y
R&Y
R
Non-latching
alarms
Alarm condition
Alarm tone on. Alarm lamp on. Alarm message. Flashing numerics.
still present.
Alarm condition
All audible and visual
no longer present.
alarm indicators
automatically stop.
Alarm condition
Alarm tone off. Alarm lamp off. Alarm message. Flashing numerics. Audible
still present.
alarm reminder (if configured).
Alarm condition
Audible and visual alarm indicators automatically stop.
no longer present.
Pause Al.
Pause Al.
5 Min.
10 Min.
R&Y
R
Off
R
Visual and audible
latching
Alarm tone on.
Alarm lamp on.
Alarm message.
Flashing numerics.
Latching Alarms
R
Off
Off
Off
Visual latching,
audible non-latching
Alarm message. Flashing
numerics.
Audible alarm indicators
automatically stop.