2 Basic Operation
2
Connect the required patient cables and sensors. The connectors are color-coded to the patient
cables and sensors for easy identification.
When connecting devices for acquiring measurements, always position cables and tubing carefully to
WARNING
avoid entanglement or potential strangulation.
Starting Monitoring
After you switch on the monitor,
Admit your patient to the monitor.
1
Check that the profile, alarm limits, alarm and QRS volumes, patient category and paced status
2
and so forth are appropriate for your patient. Change them if necessary.
Refer to the appropriate measurement section for details of how to perform the measurements you
3
require.

Disconnecting from AC Mains Power

To disconnect the monitor from AC mains power, unplug the power cord for the external power
supply (M8023A) from the mains socket.

Monitoring After a Power Failure

If external power is disconnected or there is a power failure, the monitor continues to run on its
rechargeable battery.
If the monitor is without any power (no external power or the battery is empty) for less than one
minute, monitoring will resume with all active settings unchanged. If the monitor is without power for
more than one minute, the behavior depends on your configuration. If Automat. Default is set
to Yes, the default profile will be loaded when power is restored. If Automat. Default is set to
No, all active settings are retained, if power is restored within 48 hours. The Automat. Default
setting is made in Configuration Mode.

Networked Monitoring

You can connect your monitor to an Information Center on a network, using one of the optional
interfaces:
• Standard wired LAN
• Wireless LAN
• IntelliVue Instrument Telemetry System (IIT)
Do not connect patient monitors to the standard hospital network.