25 Default Settings Appendix
SpO
Default Settings
2
SpO
Settings
2
Alarms
QRS Volume
Tone Modulation
Tone Mod. Type
Speed
Perfusion
Average
NBP Alarm Suppr.
Extd. Auto OnOff
Color
SpO
Alarm Default Settings
2
Setting
Desat Limit
Low Limit
High Limit
Desat delay
High Alarm delay
Low Alarm delay
Parameter Alarms On/Off
Label
Pulse Settings
Pulse (SpO
Pulse Alarms On/Off
Pulse High Limit
Pulse Low Limit
' Extr Brady
Brady Clamp
' ExtrTachy
Tachy Clamp
282
Adult
80
90
100
20 sec
10 sec
10 sec
on
SpO
) On/Off
on
2
on
120 bpm
50 bpm
20 bpm
40 bpm
20 bpm
200 bpm
Factory Defaults
Adult
Pedi
On
1
Yes
Enhanced
25 mm/s
On
10 sec
On
Disabled
light blue (cyan)
Pediatric
80
90
100
20 sec
10 sec
10 sec
on
SpO
2
on
on
160 bpm
75 bpm
20 bpm
40 bpm
20 bpm
220 bpm
SpO
2
Neo
Neonatal
80
85
95
20 sec
10 sec
10 sec
on
SpO
2
2
on
on
200 bpm
100 bpm
20 bpm
50 bpm
20 bpm
240 bpm
Default Settings