Philips M8102A Instructions For Use Manual Page 273

Patient monitor.
Hide thumbs

Advertisement

Measurement Specifications
CO
Alarm
2
Specifications
etCO
High
2
etCO
Low
2
imCO
2
awRR High
awRR Low
Apnea delay
Interfering Gas and Vapor Effects On CO
The specified deviations for M8102A/M3014A are valid when the appropriate corrections are switched
on and set correctly.
Gas or Vapor
Nitrous Oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Metered dose inhaler
propellants
Desflurane
Ethanol
Isopropanol
Acetone
Methane
Range
20 to 95 mmHg (2 to 13 kPa) 1 mmHg (0.1 kPa)
10 to 90 mHg (1 to 12 kPa)
High
2 to 20 mmHg
(0.3 to 3.0 kPa)
Adult/pedi: 10 to 100 rpm
Neo: 30 to 150 rpm
Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
10 to 40 seconds
Gas Level
(% volume fraction)
60
4
5
5
5
80
50
-
15
0.1
0.1
0.1
1.0
Adjustment
steps of 1 mmHg
(0.1 kPa)
under 20 rpm: 1 rpm
steps
over 20 rpm:5 rpm
steps
5 second steps
Measurement Values
2
M8102A/M3014A
Additional deviation due to gas interfence, measured at
0 - 40 mmHg CO
±1 mmHg
±2 mmHg
±2 mmHg
±2 mmHg
±2 mmHg
-5 mmHg
±1 mmHg
not specified for use
+5 mmHg
±1 mmHg
±1 mmHg
±1 mmHg
±1 mmHg
24 Specifications
Delay
M8102A/M3014A/M3016A: less than
14 seconds
M3015A: less than18 seconds.
M8102A/M3014A/M3016A: less than
14 seconds
M3015A: less than18 seconds.
M8102A/M3014A/M3016A: less than
14 seconds
M3015A: less than18 seconds.
M3015A:
settings <20 rpm: less than 8 seconds
>20 rpm: less than 18 seconds
M8102A/M3014A/M3016A
settings <20 rpm: less than 4 seconds
>20 rpm: less than 14 seconds
set apnea delay time +4 seconds
(M8102A/M3014A/M3016A) or
8 seconds (M3015A)
M3015A
2
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
265