Measurement Specifications
M3015A Microstream CO
Rise Time
Sample Flow Rate
Gas Sampling Delay Time
Sound Pressure
Total System Response Time
M3014A Mainstream CO
CO
2
awRR
Warm-up Time
Response Time
M3014A Sidestream CO
CO
2
awRR
Performance Specifications
2
190 ms for neonatal mode
(measured with FilterLine H for neonatal)
240 ms for adult mode
(measured with FilterLine H for adult)
50 + 15/-7.5 ml/minute
Maximum:
3 seconds (2m sample lines)
6 seconds (4m sample lines)
Acoustic noise: <45 dBA
The total system response time is the sum of the delay time and the rise time.
Performance Specifications
2
Range
0 to 150 mmHg (0 to 20.0 kPa)
Accuracy
after 2 minutes warmup:
For values between 0 and 40 mmHg: ±2.0 mmHg (±0.29 kPa)
For values from 41 to 70 mmHg: ±5% of reading
For values from 71 to 100 mmHg: ±8% of reading
The specifications are valid for standard gas mixtures, balance air, fully
hydrated at 35°C, P
Resolution
Numeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Stability:
Short term drift
±0.8 mmHg over four hours
Long term drift
Accuracy specification will be maintained over a 120 hour period
Range
2 to 150 rpm
Accuracy
±1 rpm
2 minutes with CO
Less than 60 ms (with adult or infant reusable or disposable adapter)
Performance Specifications
2
Range
0 to 150 mmHg (0 to 20.0 kPa)
Accuracy
after 2 minutes warmup:
For values between 0 and 40 mmHg: ±2.0 mmHg (±0.29 kPa)
For values from 41 to 70 mmHg: ±5% of reading
For values from 71 to 100 mmHg: ±8% of reading
For values from 101 to 150 mmHg: ±10% of reading
At respiration rates above 80 rpm, all ranges are ±12% of actual. The
specifications are valid for gas mixtures of CO
760 mmHg within specified operating temperature range.
Resolution
Numeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Stability:
Short term drift
±0.8 mmHg over four hours
Long term drift
Accuracy specification will be maintained over a 120 hour period
Range
2 to 150 rpm
Accuracy
±1 rpm
= 760 mmHg, flow rate = 2 l/min.
abs
transducer attached for full accuracy specification
2
, balance N
2
24 Specifications