Hide thumbs

Advertisement

M4607A Battery Specifications

Compatible Devices
IntelliVue Instrument Telemetry Wireless Network (USA only)
Internal WMTS Adapter
IntelliVue Instrument Telemetry Wireless Network (except USA)
Internal ISM Adapter
IntelliVue 802.11 Bedside Adapter (Wireless Network Adapter)
Internal Wireless Adapter
M4607A Battery Specifications
One battery is required for battery operation of the monitor.
M4607A Battery Specifications
Physical Specifications
W x D x H
Weight
Performance Specifications
Nominal Voltage
Rated Capacity at discharge C/5
Environmental Specifications
Temperature Range
Humidity Range
Battery Type
Safety
Electromagnetic Compatibility (EMC)
Communication Standard
Technology
Frequency Band
Technology
Frequency Band
Technology
Frequency Band
66 mm (2.36 in) x 80 mm (3.15 in) x 20 mm (0.79 in)
160 g ±5%
10.8 Volt
1000 mAh (typical)
Discharge 0 to 60qC (32 to 122qF)
Charge 0 to 60qC (32 to 122qF)
Storage and Transportation: -20 to 65qC (-4 to 140qF)
Operating: 15% to 95% Relative Humidity (RH)
Storage and Transportation: 5% to 95% Relative Humidity (RH)
Lithium Ion Mangan, 10.8 V, 1000 mAh,
complies with UL 2054
complies with the requirements for FCC Type B computing Device,
and EN 61000-4-2 and EN 61000-3-2
complies with the SMBus specification v1.1
compatible with Philips Cellular Telemetry System
(CTS), cellular infrastructure
WMTS, 1395-1400 MHz and 1427-1432 MHz
compatible with Philips Cellular Telemetry System
(CTS), cellular infrastructure
2.4 GHz ISM
IEEE 802.11a/b/g
2.4 GHz and 5 GHz ISM Band