Download  Print this page

Spo 2 Accessories - Philips M8102A Instructions For Use Manual

Patient monitor.
Hide thumbs

Advertisement

SpO
Accessories
2
SpO
Accessories
2
Some Nellcor sensors contain natural rubber latex which may cause allergic reactions. See the
Instructions for Use supplied with the sensors for more information. M1901B, M1902B, M1903B
and M1904B disposable sensors are not available in USA from Philips. Purchase Nellcor sensors and
adapter cables directly from Tyco Healthcare.
Do not use more than one extension cable with any sensors or adapter cables. Do not use an extension
cable with:
• Masimo adapter cables,
• Philips reusable sensors or adapter cables with part numbers ending in -L (indicates "Long" version).
All listed sensors operate without risk of exceeding 41°C on the skin, if the initial skin temperature
does not exceed 35°C.
Make sure that you use only the accessories that are specified for use with this device, otherwise patient
injury can result.
Product
Number
Philips reusable sensors.
M1191A/B
M1191AL/
BL
M1192A
M1193A
M1194A
M1195A
M1196A
Description
Adult sensor (2m cable), for patients
over 50 kg. Any finger, except thumb.
M1191A/B with longer cable (3 m)
Small adult, pediatric sensor (1.5m
cable) for patients between 15 kg and
50 kg. Any finger except thumb.
Neonatal sensor (1.5m cable) for
patients between 1kg and 4 kg. Hand
or foot.
Ear sensor (1.5m cable) for patients
more than 40 kg.
Infant sensor (1.5m cable) for patients
between 4kg and 15 kg. Any finger
except thumb.
Adult clip sensor (3m cable) for
patients over 40 kg. Any finger except
thumb.
Comments
No adapter cable
required.
No adapter cable
required.
23 Accessories
235