Functions Available When the Telemetry Data Window is Displayed 15 Assigning Two Devices to One Patient
Functions Available When the Telemetry Data
Window is Displayed
Controlling the Telemetry Device from the Bedside
You can change settings for a paired telemetry device at the monitor:
Select the telemetry data screen element
1
Select the Setup Tele pop-up key
2
The Setup Telemetry menu will appear with the settings available for the telemetry device.
These will normally include: adjusting heart rate alarms, ECG size control, selecting primary/secondary
lead, relearning arrhythmia, and selecting the arrhythmia analysis mode. The Equipment label of the
paired device is also shown here.
Viewing and Silencing Telemetry Alarms at the Bedside
When a telemetry device is paired with the monitor, telemetry alarms will also be indicated on the
monitor, in addition to the main indication at the Information Center.
If configured, a generic Tele Alarm message will appear in the alarm status area with standard
alarm tones. The Tele Alarm message will have the color and * or ! coding corresponding to the
severity of the alarm. The specific alarm message (for example *HR Low) will appear in the Telemetry
Data Window.
If configured, alarms generated from a paired telemetry device can be silenced at the bedside. Either the
monitor Silence key is configured to silence both monitor and telemetry alarms (must be
configured at the Information Center) or the Silence Bed pop-up key will be available:
Select the telemetry screen element.
1
Select the Silence Bed pop-up key.
2
Depending on your Information Center configuration, the Silence Bed key may silence both
telemetry alarms and bedside alarms.
Even when the telemetry data is not visible on the screen, you may be silencing telemetry and monitor
WARNING
alarms, if the Information Center and monitor are so configured.
Suspending Telemetry Alarms
When you select Pause Alarms or Alarms Off at the monitor, the alarms are off or paused for
the bedside measurements. When you switch alarms off or pause alarms at the Information Center,
both telemetry and monitor alarms are affected.
Refer to the Information Center Instructions for Use for more details on the Suspend/Pause alarms
behavior of the Information Center and telemetry device.
Using Standby
When you select Standby mode at the monitor, the bedside goes into Standby mode but the telemetry
device will continue monitoring.
Refer to the Information Center Instructions for Use for details on the Standby behavior of the
Information Center and telemetry device.