14 Monitoring Carbon Dioxide
To change the imCO
Select imCO
Changing the Apnea Alarm Delay
This determines the time limit after which the monitor alarms if the patient stops breathing.
In the Setup CO
1
In Setup awRR menu, select Apnea.
2
3
Choose the apnea alarm delay time.
Safety and effectiveness of the respiration measurement method in the detection of apnea, particularly
WARNING
the apnea of prematurity and apnea of infancy, has not been established.
Prolonged delay: The selected apnea alarm delay may be prolonged by up to 17 seconds, if an apnea
occurs during the automatic zero process. This applies to the Microstream (M3015A) measurement
only.
Deriving Alarms From awRR
In the Setup CO
1
2
In the Setup awRR menu, select Alarms.
3
Choose On to derive alarms from the airway respiration signal or Off to disable them.
Changing awRR Alarm Limits
In the Setup CO
1
Select High Limit to set the upper alarm limit.
2
Select Low Limit to set the lower alarm limit.
Select the appropriate setting.
3
176
high limit:
2
High and choose the high alarm limit from the pop-up list that opens.
2
menu, select awRR.
2
menu, select awRR.
2
menu, select awRR.
2
Setting up all CO
Measurements