Philips M8102A Instructions For Use Manual Page 180

Patient monitor.
Hide thumbs

Advertisement

14 Monitoring Carbon Dioxide
3
Enter the calibration value printed on the calstick then press
When the message CO
4
other cell and select Start CAL2 then press
When you see the message CO
5
Select Cal. Mode to switch calibration mode off. You cannot monitor in calibration mode.
6
Attaching and Removing the CO
Open the latch and place the
1
transducer onto the airway adapter.
Place the airway adapter in the
patient's breathing circuit between
the endotracheal tube and the Y-
piece. You may see the CO
SENSOR WARMUP message until
the transducer reaches operating
temperature. Wait until this
disappears before starting the
measurement.
To remove the transducer from the
2
airway adapter, open the latch and
pull out the airway adapter.
To prevent stress on the endotrachial tube, support the transducer and airway adapter.
WARNING
To avoid infection, use only sterilized airway adapters.
172
CAL1 done;Start CAL2 cal appears, put the transducer on the
2
cal done;leave Cal. Mode, calibration is complete.
2
Transducer
2
Airway Adapter
CO
Transducer
2
2
Measuring Mainstream CO
Confirm
.
Confirm
using M3016A
2
to start calibration.