Download  Print this page

Measuring Co 2 Using The Co 2 Option Or M3014a - Philips M8102A Instructions For Use Manual

Patient monitor.
Hide thumbs

Advertisement

Measuring CO
using the CO
2
2
Wait 2 minutes, allowing the sensor to reach its operating temperature and a stable thermal
condition.
Choose the appropriate airway adapter and connect it to the sensor
3
head. The airway adapter clicks into place when seated correctly.
To zero the sensor:
4
– expose the sensor to room air and keep it away from all sources of
CO
– in the setup menu for the CO
– When you see the message Zero done at <date and
time> on the status line, the zero calibration is finished and you
can begin monitoring.
Install the airway adapter at the proximal end of the circuit
5
between the elbow and the ventilator Y-section.
To prevent stress on the endotrachial tube, support the sensor and airway adapter.
WARNING
Position sensor cables and tubing carefully to avoid entanglement or potential strangulation. Do not
apply excessive tension to any cable.
Replace the airway adapter, if excessive moisture or secretions are observed in the tubing or if the CO
waveform changes unexpectedly without a change in patient status.
To avoid infection, use only sterilized, disinfected or disposable airway adapters.
Inspect the airway adapters prior to use. Do not use if airway adapter appears to have been damaged or
broken. Observe airway adapter color coding for patient population.
Preparing to Measure Sidestream CO
Plug the sensor cable into the monitor's CO
1
two minutes warm up time
2
Connect the cannula, airway adapter, or sample line as appropriate, to the
sensor. It will click into place when seated correctly.
3
To zero the sensor:
Option or M3014A
2
including the ventilator, the patient's breath and your own.
2
, select Start Zero Cal.
2
2
input connector. Allow the sensor
2
14 Monitoring Carbon Dioxide