Download  Print this page

Calculating Cerebral Perfusion - Philips M8102A Instructions For Use Manual

Patient monitor.
Hide thumbs

Advertisement

13 Monitoring Invasive Pressure
5
Select the calibration pressure from the list, for example 200 mmHg.
Select Confirm to recalculate the calibration factor using the applied pressure.
6
When the monitor displays <Press> calibration done at <date and time>,
7
remove the manometer tubing, syringe and extra stopcock. We recommend you replace the
transducer dome and tubing with sterile ones.
Label the transducer with the calibration factor shown in the Cal. Factor field in the
8
pressure's setup menu.
Reconnect the patient and start measuring again.
9
Troubleshooting the Pressure Calibration
The status line lists the probable cause of an unsuccessful calibration.
Message
can't calibr.; equipmt
malf.
can't calibrate;out of
range
can't calibr.; no
transd.
can't cal.; unstable
signal
can't calibr.;do zero
first

Calculating Cerebral Perfusion

The monitor can calculate the difference between mean arterial pressure and the intracranial pressure.
The difference is labeled CPP.
In the Main Setup menu, select Measurements.
1
In the Setup CPP menu, select ABP, ART, Ao, BAP or FAP as the arterial pressure source.
2
166
Corrective Action
Contact your service department. The pressure hardware is faulty.
Make sure that you have selected the value for
you are applying to the transducer, and repeat the calibration.
Make sure that the transducer is connected and try again.
Make sure there are no disturbances to the transducer, and repeat
the calibration.
No valid zero. Zero the transducer.
Calculating Cerebral Perfusion
Cal. Press
that

Advertisement

Table of Contents

This manual is also suitable for:

Intellivue mp2M3016aM3012aM3014aM3015a

Comments to this Manuals

Symbols: 0
Latest comments: