Monitoring Invasive Pressure

13
Measurements from an MMS extension connected to the monitor are not available when the monitor
WARNING
is running on battery power. They are only available when the monitor is powered by the external
power supply (M8023A).
Do not use the MP2 in combination with other monitors using an M1006A pressure module and the
CAUTION
HP1290A pressure transducer. This may cause interference on the respiration or invasive pressure
signals.
You can measure pressure using the monitor's built-in pressure measurement (optional), or one of the
MMS extensions.

Setting up the Pressure Measurement

1
Plug in the pressure cable.
Prepare the flush solution.
2
Flush the system to exhaust all air from the tubing. Ensure that the transducer and stopcocks are
3
free of air bubbles.
If air bubbles appear in the tubing system, flush the system with the infusion solution again. Air
WARNING
bubbles may lead to a wrong pressure reading.