Spo 2 Signal Quality Indicator (fast Spo 2 Only); Assessing A Suspicious Spo; Reading - Philips M8102A Instructions For Use Manual

Patient monitor.
Hide thumbs

Advertisement

SpO
Signal Quality Indicator (Fast SpO
2
Inspect the application site every two to three hours to ensure skin quality and correct optical
WARNING
alignment. If the skin quality changes, move the sensor to another site. Change the application site at
least every four hours.
CAUTION
• Injected dyes such as methylene blue, or intravascular dyshemoglobins such as methemoglobin and
carboxyhemoglobin may lead to inaccurate measurements.
• Interference can be caused by:
– High levels of ambient light or strobe lights or flashing lights (such as fire alarm lamps). (Hint:
cover application site with opaque material.)
– Electromagnetic interference.
– Excessive patient movement and vibration.
SpO
Signal Quality Indicator (Fast SpO
2
The SpO
is available) which gives an indication of the reliability of the current values.
The level to which the triangle is filled shows the quality of the signal; the indicator below shows a
medium signal quality, the signal quality is at a maximum when the triangle is completely filled.
SpO

Assessing a Suspicious SpO

Traditionally, pulse rate from SpO
the SpO
because the correct calculation of SpO
When pulse rate is very low, or strong arrhythmia is present, the SpO
the heart rate calculated from ECG but this does not indicate an inaccurate SpO
If you doubt the measured SpO
perfusion indicator instead to assess the signal quality.
N O T E
With pulse oximetry, sensor movement, ambient light (especially strobe lights or flashing lights) or
electromagnetic interference can give unexpected intermittent readings when the sensor is not attached
to a patient. Especially bandage-type sensor designs are sensitive to minimal sensor movement that
might occur when the sensor is dangling.
only)
2
numeric is displayed together with a signal quality indicator (if configured and enough space
2
2
reading. With newer algorithms, such as FAST-SpO
2
2

Reading

2
was compared with heart rate from ECG to confirm the validity of
2
is not directly linked to the correct detection of each pulse.
2
, use the signal quality indicator (if available) or the pleth wave and
10 Monitoring SpO
only)
2
, this is no longer a valid criteria
2
/Pleth pulse rate may differ from
2
value.