7 ECG, Arrhythmia, ST and QT Monitoring
You will see that
• if both the V-Tach Heart Rate Limit and the V-Tach Run Limit are exceeded, a red V-Tach alarm is
generated
• if the ventricular heart rate exceeds the V-Tach Heart Rate Limit but not the V-Tach Run Limit, a
yellow Non-Sustain VT alarm is generated.
116
** Non-Sustain VT
**Pair
PVC Run < 8
PVCs
HR > 100
PVC =
2
100
** Run PVCs
PVC Run > 2 but
HR
1
2
3
4
5
Number of Consecutive PVCs (PVC Run)
V-Tach Run Limit
***V-Tach
PVC Run
HR > 100
d
12
d
100
6
7
8
9
10
Arrhythmia Alarms
t
8
** Vent Rhythm
PVC Run > 12
d
HR
100
11
12
14
15
Ventricular Rhythm
Run Limit
V-Tach Heart
Rate Limit