Offset
O
Interfering Gas Effects:
2
N
O
2
CO
2
Halothane
Enflurane
Isoflurane
Desflurane
Sevoflurane
Acetone
Ethanol
Helium
Methane
Nitric Oxide
Oxygen Sensor, Operating Temperature
Oxygen Sensor, Expected Operating Life
CO2
Resolution
Range
Respiration Rate Measurement Range
Respiration Rate Resolution
Respiration Rate Accuracy
(All measurements made with or without a sterile jacket)
Channel
Units
Range
Resolution
Expression MR400 Instructions for Use
AGENT (Optional)
±1 percent
< 0.3 vol% @ 80 vol% N
< 0.3 vol% @ 5 vol% CO
< 0.3 vol% @ 5 vol% HAL
< 0.3 vol% @ 5 vol% ENF
< 0.3 vol% @ 5 vol% ISO
< 0.3 vol% @ 18 vol% DES
< 0.3 vol% @ 8 vol% SEV
< 0.3 vol% @ 1 vol% Acetone
< 0.3 vol% @ 0.1 vol% Ethanol
< 0.3 vol% @ 80 vol% HE
< 0.3 vol% @ 0.1 vol% Methane
< 0.3 vol% @ 50 ppm NO
15 to 35°C (59 to 95°F)
Product labeled with a use‐by date; 15 months from manufacturing
date (2500 hours at 100 percent O
12 months.
1 percent
0 to 100 percent
Bellows Respiration
0 to 60 breaths per minute
1 breath per minute
± 1 breath per minute
Temperature (Optional)
One
Celsius (°C) or Fahrenheit (°F)
20.0°C to 44.0°C (68.0°F to 111.2°F)
0.1°C (0.1°F)
O
2
2
). Exchange recommended every
2
Specifications A‐23