Download  Print this page

Setting Alarm Limits Globally - Philips Expression MR400 Instructions For Use Manual

Mri patient monitoring system.
Hide thumbs
   
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
246
247
248
249
250
251
252
253
254
255
256
257
258
259
260
261
262
263
264
265
266
267
268
269
270
271
272
273
274
275
276
277
278
279
280
281
282
283
284
285
286
287
288
289
290
291
292
293
294
295
296
297
298
299
300
301
302
303
304
305
306
307
308
309
310
311
312
313
314
315
316
317
318
319
320
321
322
323
324
325
326
327
328
329
330
331
332
333
334
335
336
337
338
339
340
341
342
343
344
345
346
347
348
349
350
351
352
353
354
355
356
357
358
359
360
361
362
363
364
365
366
367
368
369
370
371
372
373
374
375
376
377
378
379
380
381
382
383
384
385
386
387
388
389
390
391
392
393
394
395
396
397
398
399
400
401
402
403
404
405
406
407
408
409
410
411
412
413
414
415
416
417
418

Advertisement

The Extreme Tachycardia delta value will retain its adjusted value as the HR high alarm limit 
setting is adjusted downward, but will begin to shrink in value as the high alarm limit setting is 
adjusted closer to the alarm limit maximum. 
Desat Alarm Setting
The Desat alarm setting is restricted to a maximum value that is 2 less the SPO2 low alarm limit 
setting, while the minimum setting can be as low as the alarm minimum. See page 6‐11 for setting 
details.
As long as the SPO2 parameter is on, the Desat alarm value will retain this setting even if the func‐
tion is turned off and on.
The SPO2 alarm adjustments are designed to give priority to the Desat alarm setting. Therefore, 
the SPO2 low and high alarm limits may also be adjusted based on a change made to the Desat 
alarm setting. Whenever the Desat alarm setting is changed, the following rules determine the 
lower and upper SPO2 alarm limits:

Setting Alarm Limits Globally

Global changes to all of the lower and upper alarm limit settings can be made automatically by 
pressing the using the 1‐Touch Alarms key. These global changes are calculated using 
percentages selected in the 1‐Touch High % and the 1‐Touch Low % options; see page 4‐17. 
During calculations if a patient's monitored value is so high or low that it would exceed the alarm 
limit range for the parameter, then the alarm limit will be set to the highest or lowest possible 
value but not off, as indicated in the table on page 4‐22. For example, if after pressing the 1‐
Touch Alarms key, a patient's SPO2 upper alarm limit setting was 99 (the highest allowable 
value), then an alarm will be generated if the patient's reading reaches 100. (To turn off an alarm 
limit, see Setting Alarm Limits Individually on page 4‐13.)
Also note that setting alarm limits globally may result in a lower SPO2 alarm limit than the 
default setting. For a patient that has an SpO2 reading of 99%, the new upper limit will be 99 but 
the new lower limit will be 79 (instead of 85) at a 1‐Touch High % of 20 (factory default).
4‐12  Alarms                                                                                                                                     Expression MR400 Instructions for Use
If the Desat alarm was turned on after being set and turned off, then if the last set value of 
the SPO2 low alarm limit is less than the Desat alarm setting plus 2, the SPO2 low alarm 
limit will be set to the Desat alarm setting plus 2.
If the Desat alarm is adjusted upward so that the setting is greater than or equal to the 
SPO2 low alarm limit minus 2, then the SPO2 low alarm limit will be increased by the 
system such that the Desat alarm setting plus 2 is always maintained (i.e., the low alarm 
limit will always move to stay two greater than Desat alarm setting). If low alarm limit is 
altered due to this scenario, low alarm limit will retain this altered value even if the Desat 
alarm setting is adjusted downward.
The SPO2 high alarm limit will be adjusted as necessary in relation to the low alarm limit as 
per the normal behavior of low and high alarm limits.
The SPO2 low alarm limit will be allowed to be adjusted downward only until it is equal to 
the Desat alarm setting plus 2. No further downward adjustment of low alarm limit will be 
allowed until the Desat alarm setting is adjusted downward.

Advertisement

Chapters

Table of Contents

Comments to this Manuals

Symbols: 0
Latest comments: