Technical Alarms (INOPs)
Technical Alarms (Inoperative Conditions) (in alphabetic order)
Alarm Text
ECG EQUIPMENT MALF Hard
ECG LEADS OFF
<electrode> LEAD OFF
<electrode> LEAD OFF
INVALID LEADSET
NO SIGNAL
REPLACE BATTERY
2-16
Alarms
Priority
Condition
Failure of the ECG
equipment or failure to
calibrate ECG.
Hard
Multiple leads off
Hard
Primary lead is off.
If primary lead is
Note—
MCL, lead will be
identified as V in INOP
text.
Soft
Non-primary lead is off.
Hard
Bad lead selection switches
in the transceiver
Hard
Patient is out of range,
radio board has failed, ICN
connection failure, or no
batteries in transmitter.
Hard,
Dead battery. No
Latched
monitoring is occurring.
Message
remains
until
acknowle
dged by
clinician.
What to do
Replace the transceiver or
calibrate ECG with Service
Tool.
Reattach ECG leads to
patient.
Reattach ECG lead to
patient.
Reattach ECG lead to
patient.
Replace transceiver case.
Use supported leadset.
Contact Service.
Make sure that the
transceiver is in own care-
group range and has good
batteries. Replace the
transceiver if Power On Self
Test fails, and notify Service
Provider.
Replace batteries.