Calling The Calibration Menu - Dräger X-am 1100 Technical Handbook

Multi-gas
Table of Contents

Advertisement

Operation

Calling the Calibration Menu

— The calibration menu can only be
accessed by entering a password.
Password on delivery: » 001 «
— The default password on delivery can be
changed using the PC software "CC
Vision".
In measuring mode, press the
at least 5 seconds.
— The function for entering the password is
selected.
— The special symbol »
password" function) is displayed.
— The display shows » 000 « with the first
digit flashing.
Use the
key to set the flashing digit.
+
M
Press the
key, the second digit starts
OK
on
flashing.
Use the
key to set the flashing digit.
+
M
Press the
key, the third digit starts
OK
on
flashing.
Use the
key to set the flashing digit.
+
M
Press the
key to confirm the password
OK
on
once it has been set completely.
— The calibration menu functions can now
be selected by pressing the
Press the
key to call the selected
OK
on
function.
16
key for
+
M
« (for the "enter
key.
+