Reconstructed Leads

Reconstructed Leads
6-18
ECG and Arrhythmia Monitoring
Reconstruction of leads from the sourced wave is defined by the
calculations in the following table. EASI reconstructed leads are a linear
combination of all three raw EASI leads
ECG Lead
3-wire
5-wire
6-wire
Standard
I
I
I
II
II (default)
II (default) LL-RA
(default)
III
III (default)
III
(default)
-
MCL
-
aVR
aVR
-
aVL
aVL
-
aVF
aVF
-
V (default)
Va
Vb
Clinical
Calculations in
terms of electrodes
LA-RA
LL-LA
V-LA, where V=C
RA-(LA+LL)/2
LA-(RA+LL)/2
LL-(LA+RA)/2
V-(RA+LA+LL)/3, where
V=C
Va-(RA+LA+LL)/3, where
Va=V2 (default) position
Vb-(RA+LA+LL)/3, where
Vb =V5 (default) position