Technical Alarms (INOPs)
Alarm Text
TELE MALFUNCTION
Source - MX40
TELE REMOVE BATT
Source - MX40
Note —
For Information
Center Release L or M, this
INOP will appear as
"REPLACE BATTERY T".
TELE SERVICE BATT
Source - MX40
TELE WEAK SIGNAL
Source - MX40
TRANSMITTER OFF
Source - MX40
UNSUPPORTED LAN
Source - MX40
WIFI OUT OF RANGE
Source - MX40
5-20
Alarms
Priority
Condition
Hard
MX40 malfunction or
self-test failure.
Hard,
The temperature of the
Latched
lithium-ion battery is >60
C and the battery must be
removed.
Hard
The lithium-ion battery has
exceeded the maximum
charge cycle limit and
reached the end of its
useful life.
Soft
Patient is at outer range
of the radio coverage
area.
The MX40 is receiving a
weak signal with high
data loss from the AP.
Condition exists for
multiple devices in a
specific area
Hard
RF Auto Shutoff after 10
minutes of all leads off and
no SpO
sensor
2
connected.
Hard
The MX40 (WLAN) is
connected to the Access
Point, but cannot obtain an
IP address.
Hard
The MX40 (WLAN) is out
of range of an access
point.
What to do
Contact Service to replace the
MX40.
Replace the lithium-ion
o
battery.
Dispose of old battery
properly.
Replace the lithium-ion
battery.
Dispose of old battery
properly.
Return patient to the coverage
area.
If patient is in close proximity
to AP, replace the MX40.
Contact service.
The AP covering the specific
area is suspect. Contact
Service
Reattach ECG leads to
patient.
Reattach SpO
sensor.
2
Correct the IP address issue.
Correct the RF coverage issue.