Download  Print this page

Advertisement

Alarm Text
SpO
POOR SIGNAL
2
Source - MX40
SpO
T SEARCHING
2
Source - MX40
SpO
T SENSOR OFF
2
Note —
The ability of the
algorithm to detect this
condition depends on the
sensor type in use.
SpO
T SENSOR MALF
2
Source - MX40
SpO
UNKN SENSOR
2
Source - MX40
SpO
T UPGRADE
2
Source - MX40
TELE BATT EMPTY
Note —
This INOP may
also be configured to
display as a Red or Yellow
Technical Alarm.
Source - MX40
Note —
For Information
Center Release L or M, this
INOP will appear as
"REPLACE BATTERY T".
TELE BATTERY TEMP
Source - MX40
Note —
For Information
Center Release L or M, this
INOP will appear as
"REPLACE BATTERY T"
TELE CHECK BATT
Source - MX40
Priority
Condition
Soft
Although a measurement
may be possible, its
accuracy may be reduced
due to poor signal quality.
Soft
The patient signal is
analyzed, but a valid
numeric is not available
yet.
Hard
The algorithm has
determined that a sensor is
connected, but not properly
applied to the patient.
Hard
Malfunction of the SpO
sensor/adapter cable
Hard
The connected SpO
sensor and/or adapter
cable is not supported by
the hardware version.
Soft
SpO
hardware is in
2
upgrade process.
Monitoring is not possible.
Hard,
Lithium-ion battery level is
Latched
critically low. A 10-minute
countdown begins. The
MX40 will shut down if the
condition is not cleared.
Hard
The temperature of the
lithium-ion battery is above
55
o
C or below -5
Soft
Lithium-ion battery has <
25 charge cycles
remaining before reaching
the charge cycle maximum
limit.
Technical Alarms (INOPs)
What to do
Apply the sensor according to
the manufacturer's
instructions.
Relocate the sensor to a
different site on the patient.
Wait for the measurement to
complete.
Apply the sensor according to
the manufacturer's
instructions.
If the condition persists,
relocate the sensor to a
different site on the patient.
Replace sensor.
2
Use specified sensor and/or
2
adapter cable.
Wait for the upgrade process to
complete.
Insert a charged lithium-ion
battery pack.
Replace the lithium-ion battery.
o
C.
Be aware that the Lithium-ion
battery pack will soon need
replacement.
Alarms
5-19

Advertisement

Table of Contents

   Also See for Philips IntelliVue MX40

Comments to this Manuals

Symbols: 0